Strain assessment in the carotid artery wall using ultrasound speckle tracking: validation in a sheep model.
Phys Med Biol
; 60(3): 1107-23, 2015 Feb 07.
Article
en En
| MEDLINE
| ID: mdl-25586239
The aim of this study was to validate carotid artery strain assessment in-vivo using ultrasound speckle tracking. The left carotid artery of five sheep was exposed and sonomicrometry crystals were sutured onto the artery wall to obtain reference strain. Ultrasound imaging was performed at baseline and stress, followed by strain estimation using an in-house speckle tracking algorithm tuned for vascular applications. The correlation between estimated and reference strain was r = 0.95 (p < 0.001) and r = 0.87 (p < 0.01) for longitudinal and circumferential strain, respectively. Moreover, acceptable limits of agreement were found in Bland-Altman analysis (longitudinally: -0.15 to 0.42%, circumferentially: -0.54 to 0.50%), which demonstrates the feasibility of estimating carotid artery strain using ultrasound speckle tracking. However, further studies are needed to test the algorithm on human in-vivo data and to investigate its potential to detect subclinical cardiovascular disease and characterize atherosclerotic plaques.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Ultrasonido
/
Interpretación de Imagen Asistida por Computador
/
Arterias Carótidas
/
Ondas de Choque de Alta Energía
Tipo de estudio:
Diagnostic_studies
Límite:
Animals
/
Female
/
Humans
Idioma:
En
Revista:
Phys Med Biol
Año:
2015
Tipo del documento:
Article
País de afiliación:
Bélgica