Determination of Chondroitin Sulfate Content in Raw Materials and Dietary Supplements by High-Performance Liquid Chromatography with UV Detection After Enzymatic Hydrolysis: Single-Laboratory Validation First Action 2015.11.
J AOAC Int
; 99(1): 53-4, 2016.
Article
en En
| MEDLINE
| ID: mdl-26821980
A previously validated method for determination of chondroitin sulfate in raw materials and dietary supplements was submitted to the AOAC Expert Review Panel (ERP) for Stakeholder Panel on Dietary Supplements Set 1 Ingredients (Anthocyanins, Chondroitin, and PDE5 Inhibitors) for consideration of First Action Official Methods(SM) status. The ERP evaluated the single-laboratory validation results against AOAC Standard Method Performance Requirements 2014.009. With recoveries of 100.8-101.6% in raw materials and 105.4-105.8% in finished products and precision of 0.25-1.8% RSDr within-day and 1.6-4.72% RSDr overall, the ERP adopted the method for First Action Official Methods status and provided recommendations for achieving Final Action status.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Rayos Ultravioleta
/
Sulfatos de Condroitina
/
Suplementos Dietéticos
/
Enzimas
Tipo de estudio:
Diagnostic_studies
Idioma:
En
Revista:
J AOAC Int
Año:
2016
Tipo del documento:
Article
País de afiliación:
Estados Unidos