MRI-guided focused ultrasound ablation of lumbar medial branch nerve: Feasibility and safety study in a swine model.
Kaye, Elena A; Monette, Sebastien; Srimathveeravalli, Govindarajan; Maybody, Majid; Solomon, Stephen B; Gulati, Amitabh.
Afiliación
  • Kaye EA; a Department of Medical Physics , Memorial Sloan Kettering Cancer Center , New York ;
  • Monette S; b Tri-Institutional Laboratory of Comparative Pathology, Memorial Sloan Kettering Cancer Center , Rockefeller University, Weill Cornell Medical College , New York ;
  • Srimathveeravalli G; c Department of Radiology , Memorial Sloan Kettering Cancer Center , New York ;
  • Maybody M; c Department of Radiology , Memorial Sloan Kettering Cancer Center , New York ;
  • Solomon SB; c Department of Radiology , Memorial Sloan Kettering Cancer Center , New York ;
  • Gulati A; d Department of Anesthesiology , Memorial Sloan Kettering Cancer Center , New York , USA.
Int J Hyperthermia ; 32(7): 786-94, 2016 11.
Article en En | MEDLINE | ID: mdl-27443328
ABSTRACT

PURPOSE:

About 10-40% of chronic low back pain cases involve facet joints, which are commonly treated with lumbar medial branch (MB) radiofrequency neurotomy. Magnetic resonance imaging-guided focused ultrasound (MRgFUS), a non-invasive, non-ionising ablation modality used to treat tumours, neuropathic pain and painful bone metastasis can also be used to disrupt nerve conduction. This work's purpose was to study the feasibility and safety of direct MRgFUS ablation of the lumbar MB nerve in acute and subacute swine models. MATERIALS AND

METHODS:

In vivo MRgFUS ablation was performed in six swine (three acute and three subacute) using a clinical MRgFUS system and a 3-T MRI scanner combination. Behavioural assessment was performed, and imaging and histology were used to assess the treatment. RESULTS AND

CONCLUSIONS:

Histological analysis of the in vivo studies confirmed thermal necrosis of the MB nerve could be achieved without damaging the spinal cord or adjacent nerve roots. MRgFUS did not cause changes in the animals' behaviour or ambulation.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Ultrasonido Enfocado de Alta Intensidad de Ablación / Bloqueo Nervioso Límite: Animals / Female / Humans Idioma: En Revista: Int J Hyperthermia Asunto de la revista: NEOPLASIAS / TERAPEUTICA Año: 2016 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Ultrasonido Enfocado de Alta Intensidad de Ablación / Bloqueo Nervioso Límite: Animals / Female / Humans Idioma: En Revista: Int J Hyperthermia Asunto de la revista: NEOPLASIAS / TERAPEUTICA Año: 2016 Tipo del documento: Article
...