EGS_cbct: Simulation of a fan beam CT and RMI phantom for measured HU verification.
van Eeden, Dete; du Plessis, Freek.
Afiliación
  • van Eeden D; Department of Medical Physics, University of the Free State, Bloemfontein 9300, South Africa. Electronic address: detevaneeden@gmail.com.
  • du Plessis F; Department of Medical Physics, University of the Free State, Bloemfontein 9300, South Africa. Electronic address: DuPlessisFCP@ufs.ac.za.
Phys Med ; 32(10): 1375-1380, 2016 Oct.
Article en En | MEDLINE | ID: mdl-27682511
INTRODUCTION: A mathematical 3D model of an existing computed tomography (CT) scanner was created and used in the EGSnrc-based BEAMnrc and egs_cbct Monte Carlo codes. Simulated transmission dose profiles of a RMI-465 phantom were analysed to verify Hounsfield numbers against measured data obtained from the CT scanner. METHODS AND MATERIALS: The modelled CT unit is based on the design of a Toshiba Aquilion 16 LB CT scanner. As a first step, BEAMnrc simulated the X-ray tube, filters, and secondary collimation to obtain phase space data of the X-ray beam. A bowtie filter was included to create a more uniform beam intensity and to remove the beam hardening effects. In a second step the Interactive Data Language (IDL) code was used to build an EGSPHANT file that contained the RMI phantom which was used in egs_cbct simulations. After simulation a series of profiles were sampled from the detector model and the Feldkamp-Davis-Kress (FDK) algorithm was used to reconstruct transversal images. The results were tested against measured data obtained from CT scans. RESULTS: The egs_cbct code can be used for the simulation of a fan beam CT unit. The calculated bowtie filter ensured a uniform flux on the detectors. Good correlation between measured and simulated CT numbers was obtained. CONCLUSIONS: In principle, Monte Carlo codes such as egs_cbct can model a fan beam CT unit. After reconstruction, the images contained Hounsfield values comparable to measured data.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Tomografía Computarizada por Rayos X Tipo de estudio: Health_economic_evaluation Límite: Humans Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2016 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Tomografía Computarizada por Rayos X Tipo de estudio: Health_economic_evaluation Límite: Humans Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2016 Tipo del documento: Article
...