Real-time cardiac magnetic resonance cine imaging with sparse sampling and iterative reconstruction for left-ventricular measures: Comparison with gold-standard segmented steady-state free precession.
Magn Reson Imaging
; 38: 138-144, 2017 05.
Article
en En
| MEDLINE
| ID: mdl-28065694
ABSTRACT
BACKGROUND:
Segmented cine imaging with a steady-state free-precession sequence (Cine-SSFP) is currently the gold standard technique for measuring ventricular volumes and mass, but due to multi breath-hold (BH) requirements, it is prone to misalignment of consecutive slices, time consuming and dependent on respiratory capacity. Real-time cine avoids those limitations, but poor spatial and temporal resolution of conventional sequences has prevented its routine application. We sought to examine the accuracy and feasibility of a newly developed real-time sequence with aggressive under-sampling of k-space using sparse sampling and iterative reconstruction (Cine-RT).METHODS:
Stacks of short-axis cines were acquired covering both ventricles in a 1.5T system using gold standard Cine-SSFP and Cine-RT. Acquisition parameters for Cine-SSFP were acquisition matrix of 224×196, temporal resolution of 39ms, retrospective gating, with an average of 8 heartbeats per slice and 1-2 slices/BH. For Cine-RT acquisition matrix of 224×196, sparse sampling net acceleration factor of 11.3, temporal resolution of 41ms, prospective gating, real-time acquisition of 1 heart-beat/slice and all slices in one BH. LV contours were drawn at end diastole and systole to derive LV volumes and mass.RESULTS:
Forty-one consecutive patients (15 male; 41±17years) in sinus rhythm were successfully included. All images from Cine-SSFP and Cine-RT were considered to have excellent quality. Cine-RT-derived LV volumes and mass were slightly underestimated but strongly correlated with gold standard Cine-SSFP. Inter- and intra-observer analysis presented similar results between both sequences.CONCLUSIONS:
Cine-RT featuring sparse sampling and iterative reconstruction can achieve spatial and temporal resolution equivalent to Cine-SSFP, providing excellent image quality, with similar precision measurements and highly correlated and only slightly underestimated volume and mass values.Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Función Ventricular Izquierda
/
Imagen por Resonancia Cinemagnética
/
Ventrículos Cardíacos
Tipo de estudio:
Observational_studies
Límite:
Adult
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Magn Reson Imaging
Año:
2017
Tipo del documento:
Article