A gate evaluation of the sources of error in quantitative90 Y PET.
Strydhorst, Jared; Carlier, Thomas; Dieudonné, Arnaud; Conti, Maurizio; Buvat, Irène.
Afiliación
  • Strydhorst J; IMIV, U1023 Inserm/CEA/Université Paris-Sud and ERL 9218 CNRS, Université Paris-Saclay, CEA/SHFJ, Orsay 91401, France.
  • Carlier T; Department of Nuclear Medicine, Centre Hospitalier Universitaire de Nantes and CRCNA, Inserm U892, Nantes 44000, France.
  • Dieudonné A; Department of Nuclear Medicine, Hôpital Beaujon, HUPNVS, APHP and Inserm U1149, Clichy 92110, France.
  • Conti M; Siemens Healthcare Molecular Imaging, Knoxville, Tennessee, 37932.
  • Buvat I; IMIV, U1023 Inserm/CEA/Université Paris-Sud and ERL 9218 CNRS, Université Paris-Saclay, CEA/SHFJ, Orsay 91401, France.
Med Phys ; 43(10): 5320-5329, 2016 Oct.
Article en En | MEDLINE | ID: mdl-28105711
PURPOSE: Accurate reconstruction of the dose delivered by 90 Y microspheres using a postembolization PET scan would permit the establishment of more accurate dose-response relationships for treatment of hepatocellular carcinoma with 90 Y. However, the quality of the PET data obtained is compromised by several factors, including poor count statistics and a very high random fraction. This work uses Monte Carlo simulations to investigate what impact factors other than low count statistics have on the quantification of90 Y PET. METHODS: PET acquisitions of two phantoms-a NEMA PET phantom and the NEMA IEC PET body phantom-containing either 90 Y or 18 F were simulated using gate. Simulated projections were created with subsets of the simulation data allowing the contributions of random, scatter, and LSO background to be independently evaluated. The simulated projections were reconstructed using the commercial software for the simulated scanner, and the quantitative accuracy of the reconstruction and the contrast recovery of the reconstructed images were evaluated. RESULTS: The quantitative accuracy of the 90 Y reconstructions were not strongly influenced by the high random fraction present in the projection data, and the activity concentration was recovered to within 5% of the known value. The contrast recovery measured for simulated 90 Y data was slightly poorer than that for simulated 18 F data with similar count statistics. However, the degradation was not strongly linked to any particular factor. Using a more restricted energy range to reduce the random fraction in the projections had no significant effect. CONCLUSIONS: Simulations of 90 Y PET confirm that quantitative 90 Y is achievable with the same approach as that used for 18 F, and that there is likely very little margin for improvement by attempting to model aspects unique to 90 Y, such as the much higher random fraction or the presence of bremsstrahlung in the singles data.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Radioisótopos de Itrio / Tomografía de Emisión de Positrones Tipo de estudio: Evaluation_studies / Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Med Phys Año: 2016 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Radioisótopos de Itrio / Tomografía de Emisión de Positrones Tipo de estudio: Evaluation_studies / Health_economic_evaluation / Prognostic_studies Idioma: En Revista: Med Phys Año: 2016 Tipo del documento: Article País de afiliación: Francia
...