Wave-CAIPI for highly accelerated MP-RAGE imaging.
Polak, Daniel; Setsompop, Kawin; Cauley, Stephen F; Gagoski, Borjan A; Bhat, Himanshu; Maier, Florian; Bachert, Peter; Wald, Lawrence L; Bilgic, Berkin.
Afiliación
  • Polak D; Department of Radiology, A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
  • Setsompop K; Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • Cauley SF; Department of Radiology, A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
  • Gagoski BA; Harvard Medical School, Boston, Massachusetts, USA.
  • Bhat H; Harvard-MIT Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.
  • Maier F; Department of Radiology, A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
  • Bachert P; Harvard Medical School, Boston, Massachusetts, USA.
  • Wald LL; Harvard Medical School, Boston, Massachusetts, USA.
  • Bilgic B; Fetal-Neonatal Neuroimaging & Developmental Science Center, Boston Children's Hospital, Boston, Massachusetts, USA.
Magn Reson Med ; 79(1): 401-406, 2018 Jan.
Article en En | MEDLINE | ID: mdl-28220617
ABSTRACT

PURPOSE:

To introduce a highly accelerated T1-weighted magnetization-prepared rapid gradient echo (MP-RAGE) acquisition that uses wave-controlled aliasing in parallel imaging (wave-CAIPI) encoding to retain high image quality.

METHODS:

Significant acceleration of the MP-RAGE sequence is demonstrated using the wave-CAIPI technique. Here, sinusoidal waveforms are used to spread aliasing in all three directions to improve the g-factor. Combined with a rapid (2 s) coil sensitivity acquisition and data-driven trajectory calibration, we propose an online integrated acquisition-reconstruction pipeline for highly efficient MP-RAGE imaging.

RESULTS:

The 9-fold accelerated MP-RAGE acquisition can be performed in 71 s, with a maximum and average g-factor of gmax = 1.27 and gavg = 1.06 at 3T. Compared with the state-of-the-art method controlled aliasing in parallel imaging results in higher acceleration (2D-CAIPIRINHA), this is a factor of 4.6/1.4 improvement in gmax /gavg . In addition, we demonstrate a 57 s acquisition at 7T with 12-fold acceleration. This acquisition has a g-factor performance of gmax = 1.15 and gavg = 1.04.

CONCLUSION:

Wave encoding overcomes the g-factor noise amplification penalty and allows for an order of magnitude acceleration of MP-RAGE acquisitions. Magn Reson Med 79401-406, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Encéfalo / Imagen por Resonancia Magnética Límite: Female / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Encéfalo / Imagen por Resonancia Magnética Límite: Female / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos
...