Your browser doesn't support javascript.
loading
Superficial and peripheral dose in compensator-based FFF beam IMRT.
Zhang, Daniel G; Feygelman, Vladimir; Moros, Eduardo G; Latifi, Kujtim; Hoffe, Sarah; Frakes, Jessica; Zhang, Geoffrey G.
Afiliación
  • Zhang DG; Department of Chemistry, University of California, Berkeley, CA, USA.
  • Feygelman V; Radiation Oncology, Moffitt Cancer Center, Tampa, FL, USA.
  • Moros EG; Radiation Oncology, Moffitt Cancer Center, Tampa, FL, USA.
  • Latifi K; Radiation Oncology, Moffitt Cancer Center, Tampa, FL, USA.
  • Hoffe S; Radiation Oncology, Moffitt Cancer Center, Tampa, FL, USA.
  • Frakes J; Radiation Oncology, Moffitt Cancer Center, Tampa, FL, USA.
  • Zhang GG; Radiation Oncology, Moffitt Cancer Center, Tampa, FL, USA.
J Appl Clin Med Phys ; 18(1): 151-156, 2017 Jan.
Article en En | MEDLINE | ID: mdl-28291940
ABSTRACT
Flattening filter-free (FFF) beams produce higher dose rates. Combined with compensator-based intensity modulated radiotherapy (IMRT) techniques, the dose delivery for each beam can be much shorter compared to the flattened beam MLC-based or flattened beam compensator-based IMRT. This 'snap shot' IMRT delivery is beneficial to patients for tumor motion management. Due to softer energy, superficial doses in FFF beam treatment are usually higher than those from flattened beams. Due to no flattening filter, thus less photon scattering, peripheral doses are usually lower in FFF beam treatment. However, in compensator-based IMRT using FFF beams, the compensator is in the beam pathway. Does it introduce beam hardening effects and scattering such that the superficial dose is lower and peripheral dose is higher compared to FFF beam MLC-based IMRT? This study applied Monte Carlo techniques to investigate the superficial and peripheral doses in compensator-based IMRT using FFF beams and compared it to the MLC-based IMRT using FFF beams and flattened beams. Besides varying thicknesses of brass slabs to simulate varying thicknesses of compensators, a simple cone-shaped compensator was simulated to mimic a clinical application. The dose distribution in water phantom by the cone-shaped compensator was then simulated by multiple MLC-defined FFF and flattened beams with varying apertures. After normalization to the maximum dose, Dmax, the superficial and peripheral doses were compared between the FFF beam compensator-based IMRT and FFF/flattened beam MLC-based IMRT. The superficial dose at the central 0.5 mm depth was about 1% (of Dmax) lower in the compensator-based 6 MV FFF (6FFF) IMRT compared to the MLC-based 6FFF IMRT, and about 8% higher than the flattened 6 MV MLC-based IMRT dose. At 8 cm off-axis at depth of central maximum dose, dmax, the peripheral dose between the 6FFF and flattened 6 MV MLC demonstrated similar doses, while the compensator dose was about 1% (of Dmax) higher. Compensators reduce the superficial doses slightly compared to open FFF beams, but increases the peripheral doses due to scatter in the compensator.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Aceleradores de Partículas / Planificación de la Radioterapia Asistida por Computador / Fantasmas de Imagen / Radioterapia de Intensidad Modulada / Filtración / Neoplasias Tipo de estudio: Health_economic_evaluation Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Aceleradores de Partículas / Planificación de la Radioterapia Asistida por Computador / Fantasmas de Imagen / Radioterapia de Intensidad Modulada / Filtración / Neoplasias Tipo de estudio: Health_economic_evaluation Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos
...