Tibial nerve neuropathy following medial opening-wedge high tibial osteotomy-case report of a rare technical complication.
Eur J Orthop Surg Traumatol
; 27(4): 563-567, 2017 May.
Article
en En
| MEDLINE
| ID: mdl-28324204
A 63-year-old woman developed tibial nerve injury caused by an overlong K wire and 4.5-mm cortical lag screw through the first distal hole below the osteotomy during medial opening-wedge high tibial osteotomy (HTO), leading to a lack of sensation on the sole of the foot with no disturbances in motor functions. The temporary lag screw in the first distal hole below the osteotomy is often inserted by an excessive length in order to compress the potentially fractured opposite cortex. By doing so, posterior neurovascular structures including the tibial nerve and the popliteal vessels can be injured. To avoid this type of injury during medial opening-wedge HTO, proper knee position and appropriate Hohmann retractor position in combination with meticulous insertion of the K wire or screw under fluoroscopic control are essential. In addition, our study reinforces the fact that different presentations of injury to the tibial nerve should be carefully considered in the absence of common diagnostic features, including weakness of the toe flexors and posterior tibial muscle of the leg with intractable pain.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Osteotomía
/
Complicaciones Posoperatorias
/
Nervio Tibial
/
Enfermedades del Sistema Nervioso Periférico
/
Osteoartritis de la Rodilla
/
Imagenología Tridimensional
Tipo de estudio:
Observational_studies
/
Prognostic_studies
Límite:
Female
/
Humans
/
Middle aged
Idioma:
En
Revista:
Eur J Orthop Surg Traumatol
Año:
2017
Tipo del documento:
Article