Development and evaluation of a method to measure wrist range of motion on paretic hand rehabilitation device.
IEEE Int Conf Rehabil Robot ; 2017: 1337-1342, 2017 07.
Article en En | MEDLINE | ID: mdl-28814006
This paper presents the development of wrist ROM measurement system that was implemented on a rehabilitation glove for paretic hand, Easy-Flex, using a rotary potentiometer. This approach enables the device to measure the angles of up to 80° with the approximate resolution of 0.1°. The method along with a universal goniometer and a smartphone clinometer application were tested on 17 healthy participants in terms of reliability. The three instruments were able to accurately measure three different angles (flexion: -10°, neutral: 0°, and extension: 20°) with measurement errors less than 2°. Moreover, ICCs of the proposed method were close to 0.90 indicating the method implemented on Easy-Flex is reliable for measuring wrist flexion and extension measurement.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Paresia / Rehabilitación / Muñeca / Rango del Movimiento Articular Límite: Female / Humans / Male Idioma: En Revista: IEEE Int Conf Rehabil Robot Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Paresia / Rehabilitación / Muñeca / Rango del Movimiento Articular Límite: Female / Humans / Male Idioma: En Revista: IEEE Int Conf Rehabil Robot Año: 2017 Tipo del documento: Article
...