A new multipartite plate system for anterior cervical spine surgery; finite element analysis.
Simsek, Hakan; Zorlu, Emre; Kaya, Serdar; Baydogan, Murat; Atabey, Cem; Çolak, Ahmet.
Afiliación
  • Simsek H; a Department of Neurosurgery , Gülhane Military Medical Academy, Haydarpasa Teaching Hospital , Istanbul , Turkey.
  • Zorlu E; a Department of Neurosurgery , Gülhane Military Medical Academy, Haydarpasa Teaching Hospital , Istanbul , Turkey.
  • Kaya S; b Department of Neurosurgery, School of Medicine , Gülhane Military Medical Academy , Ankara , Turkey.
  • Baydogan M; c Department of Metallurgy and Materials Engineering , Istanbul Technical University , Istanbul , Turkey.
  • Atabey C; a Department of Neurosurgery , Gülhane Military Medical Academy, Haydarpasa Teaching Hospital , Istanbul , Turkey.
  • Çolak A; d Department of Neurosurgery , Central Hospital , Istanbul , Turkey.
Br J Neurosurg ; 32(3): 276-282, 2018 Jun.
Article en En | MEDLINE | ID: mdl-29256271
ABSTRACT

BACKGROUND:

There are numerous available plates, almost all of which are compact one-piece plates. During the placement of relatively long plates in the treatment of multi-level cervical pathologies, instrument related complications might appear. In order to overcome this potential problem, a novel 'articulated plate system' is designed. We aimed to delineate finite element analysis and mechanical evaluations. MATERIAL AND

METHODS:

A new plate system consisting of multi partite structure for anterior cervical stabilization was designed. Segmental plates were designed for application onto the ventral surface of the vertebral body. Plates differed from 9 to13 mm in length. There are rods at one end and hooks at the other end. Terminal points consisted of either hooks or rods at one end but the other ends are blind. Finite element and mechanical tests of the construct were performed applying bending, axial loading, and distraction forces.

RESULTS:

Finite element and mechanical testing results yielded the cut off values for functional failure and breakage of the system.

CONCLUSIONS:

The articulated system proved to be mechanically safe and it lets extension of the system on either side as needed. Ease of application needs further verification via a cadaveric study.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fusión Vertebral / Titanio / Placas Óseas / Vértebras Cervicales / Análisis de Elementos Finitos / Diseño de Equipo Límite: Humans Idioma: En Revista: Br J Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2018 Tipo del documento: Article País de afiliación: Turquía

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fusión Vertebral / Titanio / Placas Óseas / Vértebras Cervicales / Análisis de Elementos Finitos / Diseño de Equipo Límite: Humans Idioma: En Revista: Br J Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2018 Tipo del documento: Article País de afiliación: Turquía
...