Anisotropy and Spatial Heterogeneity in Quantitative Ultrasound Parameters: Relevance to the Study of the Human Cervix.
Guerrero, Quinton W; Feltovich, Helen; Rosado-Mendez, Ivan M; Carlson, Lindsey C; Li, Geng; Hall, Timothy J.
Afiliación
  • Guerrero QW; Medical Physics Department, University of Wisconsin, Madison, WI, USA.
  • Feltovich H; Medical Physics Department, University of Wisconsin, Madison, WI, USA; Maternal Fetal Medicine, Obstetrics & Gynecology, Intermountain Healthcare, Provo, UT, USA.
  • Rosado-Mendez IM; Medical Physics Department, University of Wisconsin, Madison, WI, USA.
  • Carlson LC; Medical Physics Department, University of Wisconsin, Madison, WI, USA; Maternal Fetal Medicine, Obstetrics & Gynecology, Intermountain Healthcare, Provo, UT, USA.
  • Li G; Biostatistics and Medical Informatics Department, University of Wisconsin, Madison, WI, USA.
  • Hall TJ; Medical Physics Department, University of Wisconsin, Madison, WI, USA. Electronic address: tjhall@wisc.edu.
Ultrasound Med Biol ; 44(7): 1493-1503, 2018 07.
Article en En | MEDLINE | ID: mdl-29661482
ABSTRACT
Imaging biomarkers based on quantitative ultrasound can offer valuable information about properties that inform tissue function and behavior such as microstructural organization (e.g., collagen alignment) and viscoelasticity (i.e., compliance). For example, the cervix feels softer as its microstructure remodels during pregnancy, an increase in compliance that can be objectively quantified with shear wave speed and therefore shear wave speed estimation is a potential biomarker of cervical remodeling. Other proposed biomarkers include parameters derived from the backscattered echo signal, such as attenuation and backscattered power loss, because such parameters can provide insight into tissue microstructural alignment and organization. Of these, attenuation values for the pregnant cervix have been reported, but large estimate variance reduces their clinical value. That said, parameter estimates based on the backscattered echo signal may be incorrect if assumptions they rely on, such as tissue isotropy and homogeneity, are violated. For that reason, we explored backscatter and attenuation parameters as potential biomarkers of cervical remodeling via careful investigation of the assumptions of isotropy and homogeneity in cervical tissue. Specifically, we estimated the angle- and spatial-dependence of parameters of backscattered power and acoustic attenuation in the ex vivo human cervix, using the reference phantom method and electronic steering of the ultrasound beam. We found that estimates are anisotropic and spatially heterogeneous, presumably because the tissue itself is anisotropic and heterogeneous. We conclude that appropriate interpretation of imaging biomarkers of cervical remodeling must account for tissue anisotropy and heterogeneity.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cuello del Útero / Ultrasonografía / Fantasmas de Imagen Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Female / Humans Idioma: En Revista: Ultrasound Med Biol Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cuello del Útero / Ultrasonografía / Fantasmas de Imagen Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Female / Humans Idioma: En Revista: Ultrasound Med Biol Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos
...