In Vivo Validation of Electrocardiographic Imaging.
Cluitmans, Matthijs J M; Bonizzi, Pietro; Karel, Joël M H; Das, Marco; Kietselaer, Bas L J H; de Jong, Monique M J; Prinzen, Frits W; Peeters, Ralf L M; Westra, Ronald L; Volders, Paul G A.
Afiliación
  • Cluitmans MJM; Department of Cardiology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Centre, Maastricht, the Netherlands; Department of Data Science and Knowledge Engineering, Maastricht University, Maastricht, the Netherlands.
  • Bonizzi P; Department of Data Science and Knowledge Engineering, Maastricht University, Maastricht, the Netherlands.
  • Karel JMH; Department of Data Science and Knowledge Engineering, Maastricht University, Maastricht, the Netherlands.
  • Das M; Department of Radiology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Centre, Maastricht, the Netherlands.
  • Kietselaer BLJH; Department of Cardiology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Centre, Maastricht, the Netherlands; Department of Radiology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Centre, Maastricht, the Netherlands.
  • de Jong MMJ; Department of Cardiology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Centre, Maastricht, the Netherlands.
  • Prinzen FW; Department of Physiology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Centre, Maastricht, the Netherlands.
  • Peeters RLM; Department of Data Science and Knowledge Engineering, Maastricht University, Maastricht, the Netherlands.
  • Westra RL; Department of Data Science and Knowledge Engineering, Maastricht University, Maastricht, the Netherlands.
  • Volders PGA; Department of Cardiology, Cardiovascular Research Institute Maastricht, Maastricht University Medical Centre, Maastricht, the Netherlands. Electronic address: p.volders@maastrichtuniversity.nl.
JACC Clin Electrophysiol ; 3(3): 232-242, 2017 03.
Article en En | MEDLINE | ID: mdl-29759517
ABSTRACT

OBJECTIVES:

The purpose of this study was to evaluate the accuracy of noninvasive reconstructions of epicardial potentials, electrograms, activation and recovery isochrones, and beat origins by simultaneously performing electrocardiographic imaging (ECGI) and invasive epicardial electrography in intact animals.

BACKGROUND:

Noninvasive imaging of electrical potentials at the epicardium, known as ECGI, is increasingly applied in patients to assess normal and abnormal cardiac electrical activity.

METHODS:

Body-surface potentials and epicardial potentials were recorded in normal anesthetized dogs. Computed tomography scanning provided a torso-heart geometry that was used to reconstruct epicardial potentials from body-surface potentials.

RESULTS:

Electrogram reconstructions attained a moderate accuracy compared with epicardial recordings (median correlation coefficient 0.71), but with considerable variation (interquartile range 0.36 to 0.86). This variation could be explained by a spatial mismatch (overall resolution was <20 mm) that was most apparent in regions with electrographic transition. More accurate derivation of activation times (Pearson R 0.82), recovery times (R 0.73), and the origin of paced beats (median error 10 mm; interquartile range 7 to 17 mm) was achieved by a spatiotemporal approach that incorporates the characteristics of the respective electrogram and neighboring electrograms. Reconstruction of beats from repeated single-site pacing showed a stable localization of origin. Cardiac motion, currently ignored in ECGI, correlates negatively with reconstruction accuracy.

CONCLUSIONS:

ECGI shows a decent median accuracy, but variability in electrogram reconstruction can be sizable. At present, therefore, clinical interpretations of ECGI should not be made on the basis of single electrograms only. Incorporating local spatiotemporal characteristics allows for accurate reconstruction of epicardial activation and recovery patterns, and beat origin localization to a 10-mm precision. Even more reliable interpretations are expected when the influences of cardiac motion are accounted for in ECGI.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Pericardio / Estimulación Cardíaca Artificial / Electrocardiografía Tipo de estudio: Prognostic_studies Límite: Animals / Humans Idioma: En Revista: JACC Clin Electrophysiol Año: 2017 Tipo del documento: Article País de afiliación: Países Bajos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Pericardio / Estimulación Cardíaca Artificial / Electrocardiografía Tipo de estudio: Prognostic_studies Límite: Animals / Humans Idioma: En Revista: JACC Clin Electrophysiol Año: 2017 Tipo del documento: Article País de afiliación: Países Bajos
...