Your browser doesn't support javascript.
loading
Applying Standard Clinical Chemistry Assay Validation to Droplet Digital PCR Quantitative Liquid Biopsy Testing.
Milosevic, Dragana; Mills, John R; Campion, Michael B; Vidal-Folch, Noemi; Voss, Jesse S; Halling, Kevin C; Highsmith, W Edward; Liu, Minetta C; Kipp, Benjamin R; Grebe, Stefan K G.
Afiliación
  • Milosevic D; Department of Laboratory Medicine and Pathology, Division of Clinical Biochemistry and Immunology, Mayo Clinic, Rochester, MN.
  • Mills JR; Department of Laboratory Medicine and Pathology, Division of Clinical Biochemistry and Immunology, Mayo Clinic, Rochester, MN.
  • Campion MB; Department of Laboratory Medicine and Pathology, Division of Laboratory Genetics and Genomics, Mayo Clinic, Rochester, MN.
  • Vidal-Folch N; Department of Laboratory Medicine and Pathology, Division of Laboratory Genetics and Genomics, Mayo Clinic, Rochester, MN.
  • Voss JS; Department of Laboratory Medicine and Pathology, Division of Laboratory Genetics and Genomics, Mayo Clinic, Rochester, MN.
  • Halling KC; Department of Laboratory Medicine and Pathology, Division of Laboratory Genetics and Genomics, Mayo Clinic, Rochester, MN.
  • Highsmith WE; Department of Laboratory Medicine and Pathology, Division of Laboratory Genetics and Genomics, Mayo Clinic, Rochester, MN.
  • Liu MC; Department of Laboratory Meidicne and Pathology, Division of Anatomic Pathology, Mayo Clinic, Rochester, MN.
  • Kipp BR; Department of Oncology, Division of Medical Oncology Mayo Clinic, Rochester, MN.
  • Grebe SKG; Department of Laboratory Medicine and Pathology, Division of Laboratory Genetics and Genomics, Mayo Clinic, Rochester, MN.
Clin Chem ; 64(12): 1732-1742, 2018 12.
Article en En | MEDLINE | ID: mdl-30237149
ABSTRACT

BACKGROUND:

Droplet digital PCR (ddPCR) is an emerging technology for quantitative cell-free DNA oncology applications. However, assay performance criteria must be established in a standardized manner to harness this potential. We reasoned that standard protocols used in clinical chemistry assay validation should be able to fill this need.

METHODS:

We validated KRAS, EGFR, and BRAF quantitative ddPCR assays based on the Clinical Laboratory Improvement Act regulations for laboratory-developed tests in clinical chemistry and the matching Clinical and Laboratory Standards Institute guidelines. This included evaluation of limit of the blank (LOB), limit of detection (LOD), limit of quantification (LOQ), intraassay and interassay imprecision, analytical range, dilution linearity, accuracy (including comparison with orthogonal platforms), reference range study, interference, and stability studies.

RESULTS:

For the ddPCR assays, the LOB was 4 mutant copies, LODs were 12 to 22 copies, and LOQs were 35 to 64 copies. The upper limit of the dynamic range was 30000 copies, and dilutions were linear down to the LOQs with good accuracy of spike recovery of Horizon reference material. Method comparisons with next-generation sequencing and an alternative ddPCR platform showed complete qualitative agreement and quantitative concordance, with slopes of 0.73 to 0.97 and R 2s of 0.83 to 0.99. No substantial interferences were discovered. Wild-type copy numbers in plasma ranged from 462 to 6169/mL in healthy individuals.

CONCLUSIONS:

Standard clinical chemistry assay validation protocols can be applied to quantitative ddPCR assays. This should facilitate comparison of the performance of different assays and allow establishment of minimal significant change thresholds in monitoring applications.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Análisis Mutacional de ADN / Química Clínica / Reacción en Cadena de la Polimerasa / Biopsia Líquida Tipo de estudio: Guideline / Qualitative_research Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Asunto de la revista: QUIMICA CLINICA Año: 2018 Tipo del documento: Article País de afiliación: Mongolia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Análisis Mutacional de ADN / Química Clínica / Reacción en Cadena de la Polimerasa / Biopsia Líquida Tipo de estudio: Guideline / Qualitative_research Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Asunto de la revista: QUIMICA CLINICA Año: 2018 Tipo del documento: Article País de afiliación: Mongolia
...