Ultrasound melted polymer sleeve for improved primary pedicle screw anchorage: A novel augmentation technique.
Spicher, Anna; Lindtner, Richard A; Zimmermann, Simon; Stofferin, Hannes; Schmoelz, Werner.
Afiliación
  • Spicher A; Department of Trauma Surgery, Medical University of Innsbruck, Innsbruck, Austria.
  • Lindtner RA; Department of Trauma Surgery, Medical University of Innsbruck, Innsbruck, Austria.
  • Zimmermann S; Nexilis AG, Liestal, Switzerland.
  • Stofferin H; Division of Clinical and Functional Anatomy, Department of Anatomy, Histology and Embryology, Medical University of Innsbruck, Innsbruck, Austria.
  • Schmoelz W; Department of Trauma Surgery, Medical University of Innsbruck, Innsbruck, Austria. Electronic address: werner.schmoelz@i-med.ac.at.
Clin Biomech (Bristol, Avon) ; 63: 16-20, 2019 03.
Article en En | MEDLINE | ID: mdl-30784786
ABSTRACT

BACKGROUND:

Cement augmentation of pedicle screws to prevent screw loosening is associated with significant complications, such as cement leakage or bone necrosis. Therefore, an alternative strategy to improve pedicle screw anchorage has been recently developed Polymer reinforcement of pedicle screws uses an in situ melted polymer sleeve in order to enhance screw anchorage. This biomechanical study evaluated the effect of polymer-reinforcement by comparing polymer-reinforced pedicle screws to non-augmented as well as cement-augmented screws under cyclic loading.

METHODS:

For each of the two comparisons (polymer-reinforced vs. non-augmented screws and polymer-reinforced vs. cement-augmented screws), polymer-reinforced screws and control screws were placed into the left and right pedicle of seven vertebrae (mean age 74.0 (SD 9.3) years) to allow for pairwise left-right comparisons. Each screw was subjected to cyclic cranio-caudal loading with an initial load ranging from -50 N to +50 N and with stepwise increasing compressive loads (5 N every 100 cycles) until screw loosening.

FINDINGS:

Polymer-reinforced pedicle screws resisted a higher number of load cycles until loosening than the contralateral non-augmented control screws (4300 SD 2018 vs. 2457 SD 1116 load cycles, p = 0.015). Screw anchorage of polymer-reinforced pedicle screws was comparable to that of cement augmented control screws (3857 (SD2085) vs. 4300 (SD1257) load cycles until failure, p = 0.64).

INTERPRETATION:

Our findings indicate that polymer-reinforcement significantly enhances pedicle screw anchorage in low quality bone and that its effect is similar in size than that of cement augmentation.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Polímeros / Cementos para Huesos / Tornillos Pediculares / Vértebras Lumbares Tipo de estudio: Diagnostic_studies / Etiology_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Biomech (Bristol, Avon) Asunto de la revista: ENGENHARIA BIOMEDICA / FISIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Austria

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Polímeros / Cementos para Huesos / Tornillos Pediculares / Vértebras Lumbares Tipo de estudio: Diagnostic_studies / Etiology_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Biomech (Bristol, Avon) Asunto de la revista: ENGENHARIA BIOMEDICA / FISIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Austria
...