Semi-quantification of the minimum detectable difference of imaging quality of gamma camera SPET for four radionuclides via an innovative PMMA phantom with a V-shaped slit: interpretation of a feasibility study.
Chang, Chih-Feng; Chen, Chih-Feng; Wang, Tzu-Hwei; Chiang, Fu-Tsai; Wu, Hao-Ting; Pan, Lung-Fa; Pan, Lung-Kwang.
Afiliación
  • Chang CF; Department of Cardiology, Taichung Armed Forces General Hospital, Taichung 406, Taiwan, ROC. lkpan@ctust.edu.tw, lung-fa@803.org.tw.
Hell J Nucl Med ; 22(1): 49-57, 2019.
Article en En | MEDLINE | ID: mdl-30843010
ABSTRACT

OBJECTIVE:

An indigenous polymethyl metacrylate (PMMA) phantom with a V-shaped slit and a correlated technique for semi-quantifying the minimum detectable difference (MDD) of single photon emission tomography (SPET) via gamma camera scanning are proposed and validated using four radionuclides. MATERIALS AND

METHODS:

Radio-actinide solutions of gallium-67 (67Ga), technetium-99m (99mTc), iodine-131 (131I) and thallium-201 (201Tl) were diluted to 11c.c. and thoroughly injected into the continuous zig zag slit of the PMMA phantom. Either depth or edge of the slit between two lines of the V-shape was customized from deep or wide to change into shallow or narrow gradually. Thus, the quantified MDD could be easily evaluated, according to the revised Student's t-test evaluation. The revised Student's t-test was calculated by both full width at half maximum (FWHM) and edge width between two adjacent peaks that were acquired from the original data matrix of SPET. The derived MDD was indicated as for radionuclide, depth, width in mm For 67Ga, 2.9, 2.13, for 99mTc, 2.5, 0.66, for 131I, 4.7, 2.38 and for 201Tl, 3.3, 2.00, respectively.

RESULTS:

Technetium-99m had the highest and 131I had the lowest MDD among the four radionuclides. Furthermore, two adjacent peaks of 67Ga could be easily identified with fewer counts than for 201Tl (depth, 2.9 vs. 3.3mm), but its MDD was poorer (width 2.13 vs.2.00mm). The revised Student's t-test analysis proved to be an acceptable technique for the MDD identification.

CONCLUSION:

The proposed new combination of PMMA phantom with a V-slit and the revised Student's t-test proved to be instrumental in the MDD of SPET optimization analysis.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Fantasmas de Imagen / Radiofármacos / Límite de Detección Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: Hell J Nucl Med Asunto de la revista: MEDICINA NUCLEAR Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Fantasmas de Imagen / Radiofármacos / Límite de Detección Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: Hell J Nucl Med Asunto de la revista: MEDICINA NUCLEAR Año: 2019 Tipo del documento: Article
...