Your browser doesn't support javascript.
loading
Analytical Performance of a New Immunoturbidimetric Assay for von Willebrand Factor (VWF) Activity Testing.
Clin Lab ; 65(4)2019 Apr 01.
Article en En | MEDLINE | ID: mdl-30969088
ABSTRACT

BACKGROUND:

Von Willebrand disease requires laboratory confirmation with quantitative and qualitative measurements of von Willebrand factor (VWF). Qualitative VWF-activity (VWF-Ac) tests have poor inter- and intra-laboratory reproducibility with coefficients of variation (CVs) as high as 64%, often lacking accuracy at low VWF-Ac levels.

METHODS:

This study evaluated the recently launched immunoturbidometric STAGO® STA-VWFRCo® reagent for VWF-Ac. Accuracy was evaluated on 32 samples by comparing results using the Siemens® Innovance® reagent. An intra-run reproducibility study was performed on controls. Linearity and lower limit of detection was studied on external-quality-assurance (EQA) material with a known VWF-Ac level.

RESULTS:

STA-VWFRCo® reagent results were within clinical interpretation agreement with Siemens® Innovance®. The reproducibility study yielded % CVs of 8.41 for normal and 11.46 for abnormal controls and the assay was linear between 73 and 14.6% and remained linear to 2% with extrapolation.

CONCLUSIONS:

The STAGO® STA-VWFRCo® reagent showed clinically meaningful accuracy and acceptable precision.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Enfermedades de von Willebrand / Pruebas de Coagulación Sanguínea / Factor de von Willebrand / Técnicas de Laboratorio Clínico / Inmunoturbidimetría Tipo de estudio: Diagnostic_studies / Prognostic_studies / Qualitative_research Límite: Female / Humans / Male Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Enfermedades de von Willebrand / Pruebas de Coagulación Sanguínea / Factor de von Willebrand / Técnicas de Laboratorio Clínico / Inmunoturbidimetría Tipo de estudio: Diagnostic_studies / Prognostic_studies / Qualitative_research Límite: Female / Humans / Male Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2019 Tipo del documento: Article
...