Small field correction factors determination for several active detectors using a Monte Carlo method in the Elekta Axesse linac equipped with circular cones.
Girardi, Andrea; Fiandra, Christian; Giglioli, Francesca Romana; Gallio, Elena; Ali, Omar Hammad; Ragona, Riccardo.
Afiliación
  • Girardi A; Department of Radiotherapy, Universitair Ziekenhuis Brussel, Vrije Universiteit Brussel, Brussels, Belgium. Author to whom any correspondence should be addressed.
Phys Med Biol ; 64(11): 11NT01, 2019 06 05.
Article en En | MEDLINE | ID: mdl-31051491
ABSTRACT
A Monte Carlo (MC) method was used to determine small field output correction factors for several active detectors (Exradin A16, Exradin A26, PTW microLion, PTW microDiamond, Exradin W1 and IBA RAZOR) for an Elekta Axesse linac equipped with circular cones. MC model of the linac was built with the GamBet software, using the Penelope code system. The dose-to-water simulation for each cone, ranging from 5 to 30 mm of diameter size, was used to calculate field factors and the results were validated together with Gafchromic EBT3 film. Output factors (OFs) were measured with the active detectors and correction factors were determined using the MC results. The MC simulations agreed with films within 1.2%. OFs measured with Exradin W1 scintillator were in agreement within 0.8% with MC simulations. The Exradin A16 and A26 microchambers under-responded for small fields relative to the MC (-13.1% and -4.6%, respectively). PTW microLion, IBA RAZOR and PTW microDiamond overestimated the output factor for the smallest field (+3.9%, +5.4 and +7.1%, respectively). The present study pointed out that it is crucial to apply the appropriate correction factors in order to provide accurate measurements in small beams geometry. The results showed that the Exradin W1 can be used for very small field dosimetry without correction factors, which shall be contrariwise employed for other detectors.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Aceleradores de Partículas / Radiometría / Método de Montecarlo Tipo de estudio: Health_economic_evaluation / Prognostic_studies Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Contexto en salud: 1_ASSA2030 Problema de salud: 1_financiamento_saude Asunto principal: Aceleradores de Partículas / Radiometría / Método de Montecarlo Tipo de estudio: Health_economic_evaluation / Prognostic_studies Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2019 Tipo del documento: Article
...