Characterization of high spatial resolution lithium fluoride X-ray detectors.
Mabey, P; Albertazzi, B; Michel, Th; Rigon, G; Makarov, S; Ozaki, N; Matsuoka, T; Pikuz, S; Pikuz, T; Koenig, M.
Afiliación
  • Mabey P; LULI-CNRS, Ecole Polytechnique, CEA, Université Paris-Saclay, F-91128 Palaiseau Cedex, France.
  • Albertazzi B; LULI-CNRS, Ecole Polytechnique, CEA, Université Paris-Saclay, F-91128 Palaiseau Cedex, France.
  • Michel T; LULI-CNRS, Ecole Polytechnique, CEA, Université Paris-Saclay, F-91128 Palaiseau Cedex, France.
  • Rigon G; LULI-CNRS, Ecole Polytechnique, CEA, Université Paris-Saclay, F-91128 Palaiseau Cedex, France.
  • Makarov S; Joint Institute for High Temperature RAS, Moscow 125412, Russia.
  • Ozaki N; Graduate School of Engineering, Osaka University, Suita, Osaka 565-0871, Japan.
  • Matsuoka T; Institute for Open and Transdisciplinary Research Initiatives, Osaka University, Suita, Osaka 565-0871, Japan.
  • Pikuz S; Joint Institute for High Temperature RAS, Moscow 125412, Russia.
  • Pikuz T; Joint Institute for High Temperature RAS, Moscow 125412, Russia.
  • Koenig M; LULI-CNRS, Ecole Polytechnique, CEA, Université Paris-Saclay, F-91128 Palaiseau Cedex, France.
Rev Sci Instrum ; 90(6): 063702, 2019 Jun.
Article en En | MEDLINE | ID: mdl-31255030
ABSTRACT
The response of lithium fluoride (LiF) crystal detectors to monochromatic X-rays is measured in the multi-kilo-electron-volt range. This response, as a function of the X-ray dose, is independent of photon energy with no saturation level found. The response, as a function of the incident energy flux, is found to increase for photons of lower energy due to the differing attenuation lengths of X-ray photons within the crystal. Small differences are seen between different confocal microscopes used to scan the data, suggesting the need for absolute calibration. The spatial resolution of the LiF is also measured (1.19-1.36 µm) and is found to be independent of incident photon energy. Finally, a photometric study is performed in order to assess the feasibility of using these detectors at current X-ray free electron laser and laser facilities worldwide.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2019 Tipo del documento: Article País de afiliación: Francia