Your browser doesn't support javascript.
loading
Influence of bone quality on the mechanical interaction between implant and bone: A finite element analysis.
Azcarate-Velázquez, Francisco; Castillo-Oyagüe, Raquel; Oliveros-López, Luis-Guillermo; Torres-Lagares, Daniel; Martínez-González, Álvaro-José; Pérez-Velasco, Andrea; Lynch, Christopher D; Gutiérrez-Pérez, José-Luis; Serrera-Figallo, María-Ángeles.
Afiliación
  • Azcarate-Velázquez F; Faculty of Dentistry, University of Seville, U.S., Calle Avicena s/n, 41009, Seville, Spain. Electronic address: fco_azc_vel@hotmail.com.
  • Castillo-Oyagüe R; Faculty of Dentistry, Complutense University of Madrid, U.C.M., Pza. Ramón y Cajal s/n, 28040, Madrid, Spain. Electronic address: rcastill@ucm.es.
  • Oliveros-López LG; Faculty of Dentistry, University of Seville, U.S., Calle Avicena s/n, 41009, Seville, Spain. Electronic address: lg_kn7@hotmail.com.
  • Torres-Lagares D; Faculty of Dentistry, University of Seville, U.S., Calle Avicena s/n, 41009, Seville, Spain. Electronic address: danieltl@us.es.
  • Martínez-González ÁJ; ICEMM S.L.U., 8. Oficina 0-08. Edificio Antares, Calle las Fábricas, 28923, Alcorcón, Madrid, Spain. Electronic address: alvaro.martinez@icemm.es.
  • Pérez-Velasco A; ICEMM S.L.U., 8. Oficina 0-08. Edificio Antares, Calle las Fábricas, 28923, Alcorcón, Madrid, Spain. Electronic address: andrea.perez@icemm.es.
  • Lynch CD; University Dental School & Hospital/ University College Cork, T12 E8YV, Wilton, Cork, Ireland. Electronic address: chris.lynch@ucc.ie.
  • Gutiérrez-Pérez JL; Faculty of Dentistry, University of Seville, U.S., Calle Avicena s/n, 41009, Seville, Spain. Electronic address: jlgp@us.es.
  • Serrera-Figallo MÁ; Faculty of Dentistry, University of Seville, U.S., Calle Avicena s/n, 41009, Seville, Spain. Electronic address: maserrera@us.es.
J Dent ; 88: 103161, 2019 09.
Article en En | MEDLINE | ID: mdl-31255639
ABSTRACT

OBJECTIVES:

To evaluate the influence of bone type in terms of bone density and cortical bone thickness, on the stresses induced by two implants under compressive and oblique loads.

METHODS:

A numerical simulation technique based on the finite element method was applied. Two implant types (M-12 and Astra Tech) were introduced in a model matrix whose geometry was extracted from a real CBCT radiograph of the posterior mandibular region. The Young's module and Poisson's coefficient of the bone qualities described by Misch were calculated. Loads with amplitude of 400 N were exerted in two directions compressive and 15° oblique to 5 mm above the uppermost part of the implant.

RESULTS:

The von Misses variant was analysed. Both implant types presented greater tension in the ​​cortical bone area than in the ​​trabecular bone region under compressive loading. For the oblique load condition, the stresses obtained in the cortical zone were significantly higher than those registered as a consequence of compressive loads in both implant types.

CONCLUSIONS:

Regardless of bone type, the M-12 implants presented lower tensions in the cortical bone than did the Astra implants. The tensions recorded for D3 and D4 bone types in the trabecular zone surrounding the M-12 implants were greater than those recorded for the Astra implants. CLINICAL

SIGNIFICANCE:

For both compressive and oblique loads, good mechanical behaviour was observed. The decrease in bone quality determines a worse stress distribution, and the cortical bone is overloaded. An efficient distribution of the forces may increase the implants' longevity.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Implantes Dentales / Prótesis Dental de Soporte Implantado / Análisis del Estrés Dental / Mandíbula Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Revista: J Dent Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Implantes Dentales / Prótesis Dental de Soporte Implantado / Análisis del Estrés Dental / Mandíbula Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Revista: J Dent Año: 2019 Tipo del documento: Article
...