Your browser doesn't support javascript.
loading
Interlaboratory Analysis of Isavuconazole Plasma Concentration Assays Among European Laboratories.
Pea, Federico; Krause, Robert; Müller, Carsten; Hennart, Benjamin; Richardson, Malcolm; Meinitzer, Andreas; Wiesen, Martin H J; Wiktorowicz, Tatiana; Spickermann, Jochen; Henriksen, Anne S.
Afiliación
  • Pea F; Department of Medicine, University of Udine, and Institute of Clinical Pharmacology, Santa Maria della Misericordia University Hospital, ASUIUD, Udine, Italy.
  • Krause R; Section of Infectious Diseases and Tropical Medicine, Department of Internal Medicine, Medical University of Graz, Graz, Austria.
  • Müller C; Department of Therapeutic Drug Monitoring, Center of Pharmacology, University Hospital of Cologne, Cologne, Germany.
  • Hennart B; Toxicology and Genopathy Unit, CHU Lille, Lille, France.
  • Richardson M; Mycology Reference Centre Manchester, ECMM Excellence Centre of Medical Mycology, Manchester University NHS Foundation Trust, Manchester, United Kingdom.
  • Meinitzer A; Section of Infectious Diseases and Tropical Medicine, Department of Internal Medicine, Medical University of Graz, Graz, Austria.
  • Wiesen MHJ; Department of Therapeutic Drug Monitoring, Center of Pharmacology, University Hospital of Cologne, Cologne, Germany.
Ther Drug Monit ; 41(5): 657-664, 2019 10.
Article en En | MEDLINE | ID: mdl-31568234
ABSTRACT

BACKGROUND:

Under certain circumstances, clinicians treating patients with isavuconazole for invasive aspergillosis or mucormycosis may use therapeutic drug monitoring. However, the accuracy and reproducibility of the various assays used by different laboratories for the quantification of isavuconazole plasma concentrations have yet to be determined.

METHODS:

Human plasma samples spiked with known concentrations of isavuconazole were provided to 27 European laboratories that took part in a "round-robin" test (an interlaboratory test performed independently at least 2 times; 2 rounds performed in the current study). Assay methods included liquid chromatography-tandem mass spectrometry (LC-MS/MS), LC with ultraviolet detection (LC-UV), LC with fluorescence detection (LC-FL), and bioassay. The accuracy and reproducibility compared with the known concentrations for each sample in each round were compared overall, between assays, and between laboratories.

RESULTS:

Twenty-seven laboratories participated in the study (LC-MS/MS, n = 15; LC-UV; n = 9; LC-FL, n = 1; bioassay, n = 2). In round 1, for nominal concentrations of 1000, 1700, 2500, and 4000 ng/mL, the mean (SD) determined concentrations were 1007 (183), 1710 (323), 2528 (540), and 3898 (842) ng/mL, respectively. In round 2, for nominal concentrations of 1200, 1800, 2400, and 4000 ng/mL, the mean (SD) determined concentrations were 1411 (303), 2111 (409), 2789 (511), and 4723 (798) ng/mL, respectively. Over both rounds, determined concentrations were consistently within 15% of the nominal concentrations for 10 laboratories (LC-MS/MS, n = 4; LC-UV, n = 5; bioassay, n = 1) and consistently exceeded the upper 15% margin for 7 laboratories (LC-MS/MS and LC-UV, n = 3 each; LC-FL, n = 1).

CONCLUSIONS:

Alignment of methodologies among laboratories may be warranted to improve the accuracy and reproducibility of therapeutic drug measurements.