Direct measurement of electrostatic potentials at the atomic scale: A conceptual comparison between electron holography and scanning transmission electron microscopy.
Ultramicroscopy
; 210: 112926, 2020 Mar.
Article
en En
| MEDLINE
| ID: mdl-31955112
Off-axis electron holography and first moment STEM are sensitive to electromagnetic potentials or fields, respectively. In this work, we investigate in what sense the results obtained from both techniques are equivalent and work out the major differences. The analysis is focused on electrostatic (Coulomb) potentials at atomic spatial resolution. It is shown that the probe-forming/objective aperture strongly affects the reconstructed electrostatic potentials and that, as a result of the different illumination setups, dynamical diffraction effects show a specific response with increasing specimen thickness. It is shown that thermal diffuse scattering is negligible for a wide range of specimen thicknesses, when evaluating the first moment of diffraction patterns.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Idioma:
En
Revista:
Ultramicroscopy
Año:
2020
Tipo del documento:
Article