Your browser doesn't support javascript.
loading
The Lefèvre retentive cup compared with the dual mobility cup in total hip arthroplasty revision for dislocation.
Labban, Julien; Letissier, Hoël; Mertl, Patrice; Lefèvre, Christian; Migaud, Henri; Clavé, Arnaud.
Afiliación
  • Labban J; Department of Orthopaedic and Traumatology Surgery, Brest University Hospital, Brest, France. dr.julien.labban@gmail.com.
  • Letissier H; School of Medicine, Western Brittany University, Brest, France. dr.julien.labban@gmail.com.
  • Mertl P; Department of Orthopaedic and Traumatology Surgery, Brest University Hospital, Brest, France.
  • Lefèvre C; School of Medicine, Western Brittany University, Brest, France.
  • Migaud H; Department of Orthopaedic Surgery, Amiens University Hospital, Amiens, France.
  • Clavé A; Department of Orthopaedic and Traumatology Surgery, Brest University Hospital, Brest, France.
Int Orthop ; 44(9): 1661-1667, 2020 09.
Article en En | MEDLINE | ID: mdl-32385555
ABSTRACT

BACKGROUND:

Limiting the risk of dislocation is one of the main aims of both dual mobility and Lefèvre retentive cups. However, these devices have never been compared. The goal of our study was to compare these devices in total hip arthroplasty revisions for instability. The judgement criterion was non-recurrence of dislocation in a follow-up period of eight years.

METHODS:

This retrospective case-control study compared two continuous paired series of total hip arthroplasty revisions for instability. These series included 63 patients and 159 patients with implantation of a Lefèvre retentive cup and a dual mobility cup, respectively.

RESULTS:

The success rate at eight years (i.e., no recurrence) was 91 ± 0.05% and 95 ± 0.02% in the Lefèvre retentive cup and dual mobility groups, respectively. The difference was not statistically significant (p = 0.6).

CONCLUSION:

It seems that the Lefèvre retentive cup provides comparable outcomes with the dual mobility cup in the total hip arthroplasty revisions for instability, avoiding recurrence in long term.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Luxación de la Cadera / Prótesis de Cadera Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Int Orthop Año: 2020 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Luxación de la Cadera / Prótesis de Cadera Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Int Orthop Año: 2020 Tipo del documento: Article País de afiliación: Francia
...