Pre-launch radiometric calibration of the infrared spectrometer onboard SuperCam for the Mars2020 rover.
Royer, Clément; Poulet, F; Reess, J-M; Pilorget, C; Hamm, V; Fouchet, T; Maurice, S; Forni, O; Bernardi, P; Montmessin, F; Lapauw, L; Parisot, J; Bonafous, M; Gasnault, O; Wiens, R C.
Afiliación
  • Royer C; Université Paris-Saclay, CNRS, Institut d'Astrophysique Spatiale, 91405 Orsay, France.
  • Poulet F; Université Paris-Saclay, CNRS, Institut d'Astrophysique Spatiale, 91405 Orsay, France.
  • Reess JM; LESIA, Observatoire de Paris, Université PSL, CNRS, Sorbonne Université, Université de Paris, 5 Place Jules Janssen, 92195 Meudon, France.
  • Pilorget C; Université Paris-Saclay, CNRS, Institut d'Astrophysique Spatiale, 91405 Orsay, France.
  • Hamm V; Université Paris-Saclay, CNRS, Institut d'Astrophysique Spatiale, 91405 Orsay, France.
  • Fouchet T; LESIA, Observatoire de Paris, Université PSL, CNRS, Sorbonne Université, Université de Paris, 5 Place Jules Janssen, 92195 Meudon, France.
  • Maurice S; Institut de Recherche en Astrophysique et Planétologie, CNRS, UMR 5277, 31400 Toulouse, France.
  • Forni O; Institut de Recherche en Astrophysique et Planétologie, CNRS, UMR 5277, 31400 Toulouse, France.
  • Bernardi P; LESIA, Observatoire de Paris, Université PSL, CNRS, Sorbonne Université, Université de Paris, 5 Place Jules Janssen, 92195 Meudon, France.
  • Montmessin F; LATMOS/IPSL, UVSQ Université Paris-Saclay, Sorbonne Université, CNRS, Guyancourt, France.
  • Lapauw L; LATMOS/IPSL, UVSQ Université Paris-Saclay, Sorbonne Université, CNRS, Guyancourt, France.
  • Parisot J; LESIA, Observatoire de Paris, Université PSL, CNRS, Sorbonne Université, Université de Paris, 5 Place Jules Janssen, 92195 Meudon, France.
  • Bonafous M; LESIA, Observatoire de Paris, Université PSL, CNRS, Sorbonne Université, Université de Paris, 5 Place Jules Janssen, 92195 Meudon, France.
  • Gasnault O; Institut de Recherche en Astrophysique et Planétologie, CNRS, UMR 5277, 31400 Toulouse, France.
  • Wiens RC; Los Alamos National Laboratory, Los Alamos, New Mexico 87545, USA.
Rev Sci Instrum ; 91(6): 063105, 2020 Jun 01.
Article en En | MEDLINE | ID: mdl-32611063
ABSTRACT
Near-infrared spectroscopy has become a well-known remote sensing technique for the surface characterization of planetary objects. Among them, Mars was observed in the past by three imaging spectrometers from orbit. The Infrared Spectrometer/SuperCam instrument performs near-infrared spectroscopy from the martian surface for the first time, with a 1.15 mrad field of view, in the 1.3 µm-2.6 µm range, enabling the identification of a variety of mafic and altered minerals. Before integration aboard the rover, the spectrometer underwent a calibration campaign. Here, we report the radiometric and linearity responses of the instrument, including the optical and thermal setups used to perform them over its nominal range of operations, in terms of instrument detector temperatures and spectral range. These responses were constrained by accuracy requirements (20% in absolute radiometry, 1% in relative). The derived instrument transfer function fits within these requirements (<15% in absolute and <0.8% in relative) and shall be used to calculate the expected instrumental signal-to-noise ratio for typical observation scenarios of mineral mixtures expected to be found in the Jezero crater, and ultimately to retrieve the spectral properties of the regions of interest observed by the rover.