Your browser doesn't support javascript.
loading
A robust Fourier-based method to measure pulse pressure variability.
Acosta, Sebastian; Ahmed, Mubbasheer; Yin, Suellen M; Brady, Ken M; Penny, Daniel J; Rusin, Craig G.
Afiliación
  • Acosta S; Department of Pediatrics-Cardiology, Baylor College of Medicine, Houston, TX, USA.
  • Ahmed M; Department of Pediatrics-Critical Care, Baylor College of Medicine, Houston, TX, USA.
  • Yin SM; Division of Cardiology at Children's Hospital of Philadelphia, PA, USA.
  • Brady KM; Division of Cardiac Anesthesia at Lurie Children's Hospital of Chicago, IL, USA.
  • Penny DJ; Department of Pediatrics-Cardiology, Baylor College of Medicine, Houston, TX, USA.
  • Rusin CG; Department of Pediatrics-Cardiology, Baylor College of Medicine, Houston, TX, USA.
Article en En | MEDLINE | ID: mdl-32719724
OBJECTIVE: To propose a new method to estimate pulse pressure variability (PPV) in the arterial blood pressure waveform. METHODS: Traditional techniques of calculating PPV using peak finding have a fundamental flaw that prevents them from accurately resolving PPV for small tidal volumes, limiting the use of PPV to only mechanical ventilated patients. The improved method described here addresses this limitation using Fourier analysis of an oscillatory signal that exhibits a time-varying modulation of its amplitude. The analysis reveals a constraint on the spectral representation that must be satisfied for any oscillatory signal that exhibits a time-varying modulation of its amplitude. This intrinsic mathematical structure is taken advantage of in order to improve the robustness of the algorithm. RESULTS: The applicability of the method is tested using synthetic data and 100 h of physiologic data collected from patients admitted to Texas Children's Hospital. SIGNIFICANCE AND CONCLUSION: The proposed method accurately recovers values of PPV at signal-to-noise ratios six times smaller than the traditional method. This is a significant advance for the potential use of PPV to recognize fluid responsiveness during low tidal volume ventilation or spontaneous breathing for which the signal-to-noise ratio is expected to be small.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Biomed Signal Process Control Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Biomed Signal Process Control Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos
...