Design and evaluation of quantitative MRI phantoms to mimic the simultaneous presence of fat, iron, and fibrosis in the liver.
Zhao, Ruiyang; Hamilton, Gavin; Brittain, Jean H; Reeder, Scott B; Hernando, Diego.
Afiliación
  • Zhao R; Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Hamilton G; Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Brittain JH; Department of Radiology, University of California-San Diego, San Diego, California, USA.
  • Reeder SB; Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Hernando D; Calimetrix LLC, Madison, Wisconsin, USA.
Magn Reson Med ; 85(2): 734-747, 2021 02.
Article en En | MEDLINE | ID: mdl-32783200
ABSTRACT

PURPOSE:

To design, construct, and evaluate quantitative MR phantoms that mimic MRI signals from the liver with simultaneous control of three parameters proton-density fat fraction (PDFF), R2∗ , and T1 . These parameters are established biomarkers of hepatic steatosis, iron overload, and fibrosis/inflammation, respectively, which can occur simultaneously in the liver.

METHODS:

Phantoms including multiple vials were constructed. Peanut oil was used to modulate PDFF, MnCl2 and iron microspheres were used to modulate R2∗ , and NiCl2 was used to modulate the T1 of water (T1,water ). Phantoms were evaluated at both 1.5 T and 3 T using stimulated-echo acquisition-mode MRS and chemical shift-encoded MRI. Stimulated-echo acquisition-mode MRS data were processed to estimate T1,water , T1,fat , R2,water∗ , and R2,fat∗ for each vial. Chemical shift-encoded MRI data were processed to generate PDFF and R2∗ maps, and measurements were obtained in each vial. Measurements were evaluated using linear regression and Bland-Altman analysis.

RESULTS:

High-quality PDFF and R2∗ maps were obtained with homogeneous values throughout each vial. High correlation was observed between imaging PDFF with target PDFF (slope = 0.94-0.97, R2 = 0.994-0.997) and imaging R2∗ with target R2∗ (slope = 0.84-0.88, R2 = 0.935-0.943) at both 1.5 T and 3 T. The values of R2,fat∗ and R2,water∗ were highly correlated with slope close to 1.0 at both 1.5 T (slope = 0.90, R2 = 0.988) and 3 T (slope = 0.99, R2 = 0.959), similar to the behavior observed in vivo. The value of T1,water (500-1200 ms) was controlled with varying NiCl2 concentration, while T1,fat (300 ms) was independent of NiCl2 concentration.

CONCLUSION:

Novel quantitative MRI phantoms that mimic the simultaneous presence of fat, iron, and fibrosis in the liver were successfully developed and validated.