Vertical cutoff methods in serum protein electrophoresis for the measurement of monoclonal protein concentrations: Which is best?
Cárdenas Fernández, M C; Pérez Surribas, D; Pérez Garay, R; Jiménez Jiménez, J; Gella Concustell, A; Zapico Muñiz, E; Cruz Iglesias, E; Fernández García, M; Hermoso Durán, S; Marcaida Benito, G; Valldecabres Ortiz, C.
Afiliación
  • Cárdenas Fernández MC; Department of Clinical Analysis, Hospital Clínico San Carlos, Madrid, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain. Electronic address: mariacruz.cardenas@salud.madrid.org.
  • Pérez Surribas D; Laboratori Pasteur, Andorra La Vella, Andorra; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Pérez Garay R; Department of Clinical Analysis, Hospital de Cruces, Barakaldo, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Jiménez Jiménez J; Department of Clinical Analysis, Hospital Universitario Severo Ochoa, Leganés, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Gella Concustell A; Institut de Neurociències, Universidad Autónoma de Barcelona, Barcelona, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Zapico Muñiz E; Department of Clinical Biochemistry, Hospital de la Santa Creu i Sant Pau, Barcelona, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Cruz Iglesias E; Department of Clinical Biochemistry, Hospital Universitario Basurto, Bilbao, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Fernández García M; Department of Clinical Analysis, Hospital Universitario Río Hortega, Valladolid, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Hermoso Durán S; Instituto de Investigación Sanitaria Aragón, Zaragoza, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Marcaida Benito G; Department of Clinical Analysis, Hospital General Universitario de Valencia, Valencia, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
  • Valldecabres Ortiz C; Department of Clinical Biochemistry, Hospital Universitario de La Ribera, Alcira, Spain; Protein Commission, Spanish Society of Laboratory Medicine (SEQC(ML)), Barcelona, Spain.
Clin Chim Acta ; 510: 573-580, 2020 Nov.
Article en En | MEDLINE | ID: mdl-32791139
ABSTRACT

BACKGROUND:

Monoclonal protein (M-protein) concentrations are measured by serum protein electrophoresis (SPE). Two methods are used for demarcating the M-protein area in the electropherogram perpendicular drop (PD) and tangent skimming (TS). The aim of this study was tocompare both methods and to establish which is the most accurate and precise.

METHODS:

We studied 24 sera containing M-protein (5-44 g/L). The systematic error (SE) was evaluated in a dilution series of 12 sera. Within-day, between-day, and interobserver variability were assessed. SPE was performed by capillary and agarose gel electrophoresis. M-protein concentrations were measured using both cutoff methods.

RESULTS:

The PD method shows a constant SE ranged 1.00-2.27 g/L, while constant SE for TS is ranged -0.30--0.57 g/L. None of the cutoff methods or electrophoretic methods showed a proportional SE, with the exception of the TS method in capillary electrophoresis for ß-migrating M-protein. The PD method was more precise than the TS method in all three estimates of imprecision. An increased CV for concentrations < 10 g/L in between-day imprecision was observed with the TS method. Interobserver imprecision was greater for M-protein concentrations < 17 g/L for both cutoff methods (14.85%, 26.42% respectively).

CONCLUSIONS:

Despite being less precise, the TS method provides a more accurate measurement of M-protein concentration.