Restoration of normal knee kinematics with respect to tibial insert design in mobile bearing lateral unicompartmental arthroplasty using computational simulation.
Koh, Yong-Gon; Lee, Jin-Ah; Chung, Hyun-Seok; Kim, Hyo-Jeong; Kang, Kyoung-Tak.
Afiliación
  • Koh YG; Joint Reconstruction Center, Department of Orthopaedic Surgery, Yonsei Sarang Hospital, Seoul, Republic of Korea.
  • Lee JA; Department of Mechanical Engineering, Yonsei University, Seoul, Republic of Korea.
  • Chung HS; Joint Reconstruction Center, Department of Orthopaedic Surgery, Yonsei Sarang Hospital, Seoul, Republic of Korea.
  • Kim HJ; Department of Sport and Healthy Aging, Korea National Sport University, Seoul, Republic of Korea.
  • Kang KT; Department of Mechanical Engineering, Yonsei University, Seoul, Republic of Korea.
Bone Joint Res ; 9(7): 421-428, 2020 Jul.
Article en En | MEDLINE | ID: mdl-32864113
ABSTRACT

AIMS:

Mobile-bearing unicompartmental knee arthroplasty (UKA) with a flat tibial plateau has not performed well in the lateral compartment, leading to a high rate of dislocation. For this reason, the Domed Lateral UKA with a biconcave bearing was developed. However, medial and lateral tibial plateaus have asymmetric anatomical geometries, with a slightly dished medial and a convex lateral plateau. Therefore, the aim of this study was to evaluate the extent at which the normal knee kinematics were restored with different tibial insert designs using computational simulation.

METHODS:

We developed three different tibial inserts having flat, conforming, and anatomy-mimetic superior surfaces, whereas the inferior surface in all was designed to be concave to prevent dislocation. Kinematics from four male subjects and one female subject were compared under deep knee bend activity.

RESULTS:

The conforming design showed significantly different kinematics in femoral rollback and internal rotation compared to that of the intact knee. The flat design showed significantly different kinematics in femoral rotation during high flexion. The anatomy-mimetic design preserved normal knee kinematics in femoral rollback and internal rotation.

CONCLUSION:

The anatomy-mimetic design in lateral mobile UKA demonstrated restoration of normal knee kinematics. Such design may allow achievement of the long sought normal knee characteristics post-lateral mobile UKA. However, further in vivo and clinical studies are required to determine whether this design can truly achieve a more normal feeling of the knee and improved patient satisfaction.Cite this article Bone Joint Res 2020;9(7)421-428.