Characterization of the stimulation output of four devices for focal muscle vibration.
Botter, Alberto; Cerone, Giacinto Luigi; Saggini, Raoul; Massazza, Giuseppe; Minetto, Marco Alessandro.
Afiliación
  • Botter A; Laboratory for Engineering of the Neuromuscular System (LISiN), Department of Electronics and Telecommunications, Politecnico di Torino, Corso Duca degli Abruzzi 24, 10129 Turin, Italy; PoliToBIOMed Lab, Politecnico di Torino, Turin, Italy. Electronic address: alberto.botter@polito.it.
  • Cerone GL; Laboratory for Engineering of the Neuromuscular System (LISiN), Department of Electronics and Telecommunications, Politecnico di Torino, Corso Duca degli Abruzzi 24, 10129 Turin, Italy; PoliToBIOMed Lab, Politecnico di Torino, Turin, Italy.
  • Saggini R; Department of Medical and Oral Science and Biotechnologies, G. d'Annunzio University, Chieti-Pescara, Chieti, Italy.
  • Massazza G; Division of Physical Medicine and Rehabilitation, Department of Surgical Sciences, University of Turin, Turin, Italy.
  • Minetto MA; Division of Physical Medicine and Rehabilitation, Department of Surgical Sciences, University of Turin, Turin, Italy.
Med Eng Phys ; 85: 97-103, 2020 11.
Article en En | MEDLINE | ID: mdl-33081969
ABSTRACT
Different devices for mechano-acoustic muscle vibration became available on the market in the last ten years. Although the use of these vibrators is increasing in research and clinical settings, the features of their stimulation output were never described in literature. In this study we aimed to quantify and compare the stimulation output of the four most widespread pneumatic devices for focal muscle vibration available on the market. A piezoelectric pressure sensor was used to measure the pressure profile generated by the four selected devices in the following experimental conditions i) measurement of the output changes associated with variations of the stimulation amplitude for three stimulation frequencies (100 Hz, 200 Hz, and 300 Hz); ii) measurement of the output changes during a 20-min long stimulation at constant frequency (300 Hz) and amplitude; iii) measurement of the output changes associated with the progressive activation of all stimulation channels at constant frequency (200 Hz) for different amplitudes. The maximum peak-to-peak amplitudes of the pressure waves were in the range 102 mbar - 369 mbar (below the maximum values declared by the different manufacturers). The shape of the pressure waves generated by the four devices was quasi-sinusoidal and asymmetric with respect to the atmospheric pressure. All output features had a remarkable intra- and inter-device variability. Further studies are required to support the technological improvement of the currently available devices and to focus the issues of vibration effectiveness, limitations, proper protocols, modalities of its application and assessment in neuromuscular training and rehabilitation.