Compressive stress in periodontal ligament under orthodontic movements during periodontal breakdown.
Moga, Radu Andrei; Buru, Stefan Marius; Chiorean, Cosmin Gruia; Cosgarea, Raluca.
Afiliación
  • Moga RA; Department of Cariology, Endodontics and Oral Pathology, Faculty of Dental Medicine, Iuliu Hatieganu University of Medicine and Pharmacy, Cluj-Napoca, Romania. Electronic address: randreimoga@umfcluj.ro.
  • Buru SM; Department of Structural Mechanics, Faculty of Civil Engineering, Technical University of Cluj-Napoca, Cluj-Napoca, Romania.
  • Chiorean CG; Department of Structural Mechanics, Faculty of Civil Engineering, Technical University of Cluj-Napoca, Cluj-Napoca, Romania.
  • Cosgarea R; Department of Periodontology & Peri-implant Diseases, Philipps University of Marburg, Marburg, Germany; Department of Periodontology, Operative and Preventive Dentistry, University of Bonn, Bonn, Germany; Department of Prosthodontics, Faculty of Dental Medicine, Iuliu Hatieganu University of Med
Am J Orthod Dentofacial Orthop ; 159(3): e291-e299, 2021 Mar.
Article en En | MEDLINE | ID: mdl-33487500
ABSTRACT

INTRODUCTION:

This analysis aimed to assess quantitatively and qualitatively the compressive stress (S3) in periodontal ligament in a gradual periodontal breakdown (0-8 mm) under orthodontic movements. Correlations between the applied forces, the level of bone resorption, the decrease of force magnitude, and S3 increase were also conducted.

METHODS:

On the basis of cone-beam computed tomography examinations (voxel size, 0.075 mm), nine 3-dimensional models of the second mandibular premolar with intact periodontium were created and then individually subjected to various levels of horizontal bone loss. Orthodontic forces (intrusion 0.2 N; extrusion, rotation, tipping 0.6 N; translation 1.2 N) were applied on the brackets. Finite elements analysis was performed, and S3 stresses were quantitatively and qualitatively determined.

RESULTS:

Translation and rotation induced the highest stress apically and cervically, whereas intrusion determined the lowest. Apical stress was lower than cervical stress. In intact periodontium, only intrusion and extrusion exhibited S3 stresses lower (apically and cervically) than maximum hydrostatic pressure (MHP) and maximum tolerable stress (MTS). In reduced periodontium, S3 stress (except for intrusion) exceeded MHP and MTS.

CONCLUSIONS:

In reduced periodontium, forces of 0.2 N seems safe to be used. Forces of 0.6-1.2 N may produce stresses exceeding both MTS and MHP, endangering the periodontium. S3 failure criterion (despite its widely use) seems not to be adequate for accurate quantitative results when evaluating the stress in the periodontal ligament while remaining adequate for qualitative results. An overall correlation between the applied force, S3 increase, and periodontal breakdown applicable to all 5 movements could not be established-this was possible only for sole movements.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ligamento Periodontal / Técnicas de Movimiento Dental Tipo de estudio: Prognostic_studies / Qualitative_research Límite: Humans Idioma: En Revista: Am J Orthod Dentofacial Orthop Asunto de la revista: ODONTOLOGIA / ORTODONTIA Año: 2021 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ligamento Periodontal / Técnicas de Movimiento Dental Tipo de estudio: Prognostic_studies / Qualitative_research Límite: Humans Idioma: En Revista: Am J Orthod Dentofacial Orthop Asunto de la revista: ODONTOLOGIA / ORTODONTIA Año: 2021 Tipo del documento: Article
...