Your browser doesn't support javascript.
loading
A novel retractor for reducing operation time and radiation exposure in percutaneous pedicle screw placement.
Niu, Lei; He, Rui; Duan, Liqun; Zhang, Wenzhi; Li, Xu; Cai, Haiping.
Afiliación
  • Niu L; The First Affiliated Hospital of University of Science and Technology of China.
Turk Neurosurg ; 2020 Oct 01.
Article en En | MEDLINE | ID: mdl-33759168
ABSTRACT

AIM:

Design a novel hand-held retractor to reduce operation time and radiation exposure in percutaneous pedicle screw placement. MATERIAL AND

METHODS:

126 patients with single segment thoracolumbar vertebral fracture were surgically treated with percutaneous pedicle screws through our novel hand-held retractor technique(group A) or conventional fluoroscopic method(group B), the operation time and fluoroscopy shot times were compared, and the accuracy of screw placement were assessed.

RESULTS:

There was no serious complications occurring during our study, such as infection, blood vessel injury, spinal cord or nerve root injury. We did not find any statistically difference between the two groups in corrected rate of regional Cobb's angle or vertebral body height percentage(P 0.05), however, the mean operation time was found to be 75.9±2.37min in the novel hand-held retractor method group and 94.2±2.19min in the conventional method group. The difference was statistically significant (p﹤0.001). Fluoroscopy shot times averaged 9.01±0.41 in the novel hand-held retractor group versus 16.8±0.56 in the conventional group (P 0.001), the novel hand-held retractor group had apparent advantages over the conventional method in postoperative improvement on visual analog scale (VAS) scores and Oswestry Disability Index (ODI) at 2days, 3month after operation and last follow-up (P 0.05). There was no statistical difference between group A and B in the radiographic results and screw position violation grade.

CONCLUSION:

The novel hand-held method retractors have several advantages, including shorter operation time, less fluoroscopy shot times, and better postoperative improvement on visual analog scale (VAS) scores and Oswestry Disability Index (ODI). It provides a new alternative method for effective management of thoracolumbar fractures.