Combined 23 Na and 13 C imaging at 3.0 Tesla using a single-tuned large FOV birdcage coil.
Kaggie, Joshua D; Lanz, Titus; McLean, Mary A; Riemer, Frank; Schulte, Rolf F; Benjamin, Arnold J V; Kessler, Dimitri A; Sun, Chang; Gilbert, Fiona J; Graves, Martin J; Gallagher, Ferdia A.
Afiliación
  • Kaggie JD; Department of Radiology, University of Cambridge, Cambridge, United Kingdom.
  • Lanz T; Cambridge University Hospitals, Addenbrooke's Hospital, Cambridge, United Kingdom.
  • McLean MA; Cancer Research UK Cambridge Centre, University of Cambridge, Cambridge, United Kingdom.
  • Riemer F; RAPID Biomedical GmbH, Rimpar, Germany.
  • Schulte RF; Department of Radiology, University of Cambridge, Cambridge, United Kingdom.
  • Benjamin AJV; Cambridge University Hospitals, Addenbrooke's Hospital, Cambridge, United Kingdom.
  • Kessler DA; Cancer Research UK Cambridge Centre, University of Cambridge, Cambridge, United Kingdom.
  • Sun C; Mohn Medical Imaging and Visualisation Centre (MMIV), Department of Radiology, Haukeland University Hospital, Bergen, Norway.
  • Gilbert FJ; GE Healthcare, Munich, Germany.
  • Graves MJ; Department of Radiology, University of Cambridge, Cambridge, United Kingdom.
  • Gallagher FA; Cambridge University Hospitals, Addenbrooke's Hospital, Cambridge, United Kingdom.
Magn Reson Med ; 86(3): 1734-1745, 2021 09.
Article en En | MEDLINE | ID: mdl-33934383
ABSTRACT

PURPOSE:

An unmet need in carbon-13 (13 C)-MRI is a transmit system that provides uniform excitation across a large FOV and can accommodate patients of wide-ranging body habitus. Due to the small difference between the resonant frequencies, sodium-23 (23 Na) coil developments can inform 13 C coil design while being simpler to assess due to the higher naturally abundant 23 Na signal. Here we present a removable 23 Na birdcage, which also allows operation as a 13 C abdominal coil.

METHODS:

We demonstrate a quadrature-driven 4-rung 23 Na birdcage coil of 50 cm in length for both 23 Na and 13 C abdominal imaging. The coil transmit efficiencies and B1+ maps were compared to a linearly driven 13 C Helmholtz-based (clamshell) coil. SNR was investigated with 23 Na and 13 C data using an 8-channel 13 C receive array within the 23 Na birdcage.

RESULTS:

The 23 Na birdcage longitudinal FOV was > 40 cm, whereas the 13 C clamshell was < 32 cm. The transmit efficiency of the birdcage at the 23 Na frequency was 0.65 µT/sqrt(W), similar to the clamshell for 13 C. However, the coefficient of variation of 23 Na- B1+ was 16%, nearly half that with the 13 C clamshell. The 8-channel 13 C receive array combined with the 23 Na birdcage coil generated a greater than twofold increase in 23 Na-SNR from the central abdomen compared with the birdcage alone.

DISCUSSION:

This 23 Na birdcage coil has a larger FOV and improved B1+ uniformity when compared to the widely used clamshell coil design while also providing similar transmit efficiency. The coil has the potential to be used for both 23 Na and 13 C imaging.