Lot variation and inter-device differences contribute to poor analytical performance of the DCA Vantage™ HbA1c POCT instrument in a true clinical setting.
Abildgaard, Anders; Knudsen, Cindy Søndersø; Bjerg, Lise Nørkjær; Lund, Sten; Støy, Julie.
Afiliación
  • Abildgaard A; Department of Clinical Biochemistry, Regional Hospital Horsens, Horsens, Denmark.
  • Knudsen CS; Steno Diabetes Center Aarhus, Aarhus University Hospital, Aarhus N, Denmark.
  • Bjerg LN; Department of Clinical Biochemistry, Aarhus University Hospital, Aarhus N, Denmark.
  • Lund S; Department of Clinical Biochemistry, Regional Hospital Central Jutland, Viborg, Denmark.
  • Støy J; Steno Diabetes Center Aarhus, Aarhus University Hospital, Aarhus N, Denmark.
Clin Chem Lab Med ; 60(1): 127-134, 2022 01 26.
Article en En | MEDLINE | ID: mdl-34480844
OBJECTIVES: The glycated haemoglobin fraction A1c (HbA1c) is widely used in the management of diabetes mellitus, and the Siemens DCA Vantage™ point-of-care testing (POCT) instrument offers rapid HbA1c results even far from a clinical laboratory. However, the analytical performance has been questioned, and not much is known about effects of changing reagent lot, instrument and operator. We therefore compared the analytical performance of the DCA Vantage™ with established routine methods (Tosoh G8/G11 ion exchange HPLC) in a true clinical setting at two Danish hospitals. METHODS: We extracted all routine clinical HbA1c results incidentally drawn from the same patient within 48 h (n=960 pairs) and evaluated the effect of reagent lot, operator and instrument. We also performed a prospective method comparison in our diabetes out-patient clinic (n=97). RESULTS: The critical difference (CD) between two POCT results varied between 5.14 and 6.61 mmol/mol (0.47-0.55%), and the analytical imprecision of the DCA Vantage™ (CVA) was >3%. Significant effect of reagent lot and inter-instrument differences were found, whereas no effect of operator was seen. CONCLUSIONS: The DCA Vantage™ HbA1c analysis does not fulfil the prevailing analytical performance specifications, but rigorous validation of new reagent lots and continuous recalibration of instruments may potentially improve the precision substantially. Our findings, therefore, clearly emphasise the necessity of a close collaboration between clinicians and laboratory professionals in the POCT field. Finally, POCT HbA1c results should always be interpreted together with other measures of glycaemic control to avoid inappropriate change of patient treatments due to measurement uncertainty.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Sistemas de Atención de Punto / Diabetes Mellitus Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2022 Tipo del documento: Article País de afiliación: Dinamarca

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Sistemas de Atención de Punto / Diabetes Mellitus Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2022 Tipo del documento: Article País de afiliación: Dinamarca
...