Magnetic resonance imaging safety in patients with abandoned or functioning epicardial pacing leads.
Vuorinen, Aino-Maija; Paakkanen, Riitta; Karvonen, Jarkko; Sinisalo, Juha; Holmström, Miia; Kivistö, Sari; Peltonen, Juha I; Kaasalainen, Touko.
Afiliación
  • Vuorinen AM; Radiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland. aino-maija.vuorinen@hus.fi.
  • Paakkanen R; Heart and Lung Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland.
  • Karvonen J; Heart and Lung Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland.
  • Sinisalo J; Heart and Lung Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland.
  • Holmström M; Radiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland.
  • Kivistö S; Radiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland.
  • Peltonen JI; Radiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland.
  • Kaasalainen T; Radiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00029 HUS, Helsinki, Finland.
Eur Radiol ; 32(6): 3830-3838, 2022 Jun.
Article en En | MEDLINE | ID: mdl-34989847
ABSTRACT

OBJECTIVES:

The European Society of Cardiology Guidelines on cardiac pacing from 2021 allow magnetic resonance imaging (MRI) in patients with cardiac implantable electronic devices (CIEDs) but do not recommend MRI in patients with epicardial pacing leads. The clinical dilemma remains whether performing an MRI in patients with CIED and epicardial leads is safe. We aimed to evaluate the safety of performing an MRI in patients with CIED and abandoned or functioning epicardial pacing leads.

METHODS:

We included all adult patients who underwent clinically indicated MRIs with CIED and functioning or abandoned epicardial leads in a single tertiary hospital between November 2011 and October 2019. The data were retrospectively collected.

RESULTS:

Twenty-six MRIs were performed on 17 patients with functioning or abandoned epicardial pacing leads. Sixty-nine percent of the MRI scans (18/26) were conducted on patients with functioning epicardial pacing leads. A definite adverse event occurred in one MRI scan. This was a transient elevation of the pacing threshold in a patient with a functioning epicardial ventricular pacing lead implanted 29 years previously. An irreversible atrial pacing lead impedance elevation was detected 6 months after the MRI in another patient; the association with the previous MRI remained unclear. No adverse events were detected in MRIs performed on patients with modern (implanted in 2000 or later) functioning epicardial leads.

CONCLUSIONS:

MRIs in patients with CIED and modern functioning epicardial pacing leads were performed without detectable adverse events. Further large-scale studies are necessary to confirm MRI safety in patients with epicardial pacing leads. KEY POINTS • Currently, MRI in patients with cardiac implantable electronic devices (CIEDs) and functioning or abandoned epicardial pacing leads is not recommended. • MRIs in patients with CIED and modern functioning epicardial leads (implanted in 2000 or later) were performed without detectable adverse events in our patient cohort. • Allowing MRI in patients with epicardial pacing leads may significantly improve the diagnostic work-up, especially in specific patient groups, such as patients with congenital heart disease.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Desfibriladores Implantables / Cardiopatías Congénitas Tipo de estudio: Guideline / Observational_studies / Risk_factors_studies Límite: Adult / Humans Idioma: En Revista: Eur Radiol Asunto de la revista: RADIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Finlandia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Desfibriladores Implantables / Cardiopatías Congénitas Tipo de estudio: Guideline / Observational_studies / Risk_factors_studies Límite: Adult / Humans Idioma: En Revista: Eur Radiol Asunto de la revista: RADIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Finlandia
...