Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.
Cheng, Shi-Ming; Zhang, Jia-Sheng; Shao, Xu; Wu, Ze-Tong; Li, Tian-Tian; Wang, Po; Lin, Jun-Hai; Yu, A-Yong.
Afiliación
  • Cheng SM; Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, 270 West Xueyuan Road, Wenzhou, Zhejiang, 325027, People's Republic of China.
  • Zhang JS; National Clinical Research Center for Ocular Diseases, Wenzhou, Zhejiang, People's Republic of China.
  • Shao X; Ophthalmology Department, Taihe Hospital, Hubei University of Medicine, Shiyan, Hubei, People's Republic of China.
  • Wu ZT; Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, 270 West Xueyuan Road, Wenzhou, Zhejiang, 325027, People's Republic of China.
  • Li TT; National Clinical Research Center for Ocular Diseases, Wenzhou, Zhejiang, People's Republic of China.
  • Wang P; Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, 270 West Xueyuan Road, Wenzhou, Zhejiang, 325027, People's Republic of China.
  • Lin JH; National Clinical Research Center for Ocular Diseases, Wenzhou, Zhejiang, People's Republic of China.
  • Yu AY; Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, 270 West Xueyuan Road, Wenzhou, Zhejiang, 325027, People's Republic of China.
Graefes Arch Clin Exp Ophthalmol ; 260(7): 2271-2281, 2022 Jul.
Article en En | MEDLINE | ID: mdl-35171331
ABSTRACT

PURPOSE:

To investigate the repeatability of Anterion and compare the agreement of ocular biometric measurements and predicted intraocular lens (IOL) powers with other three optical biometers.

METHODS:

Flat keratometry (Kf), steep keratometry (Ks), J0 and 45 vectors, central cornea thickness (CCT), anterior chamber depth (ACD), and axial length (AL) from the Anterion, IOLMaster 700, Lenstar LS 900, and OA-2000 were recorded. The IOL powers were calculated with the Hoffer Q, Holladay 1, SRK/T, and Haigis formulas. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability coefficient (RC), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Inter-device agreement between the four biometers was assessed with the 95% limits of agreement.

RESULTS:

In total, 101 right eyes of 101 participants were enrolled. The Anterion showed good repeatability for all the included biometric parameters with all the CoV ≤ 0.30% and ICC ≥ 0.930 except for J45 with moderate repeatability (ICC was 0.849). Good agreement was found among the four devices for Kf, Ks, J0, J45, ACD, and AL. Generally, wide 95% LoA was found for the predicted IOL powers with the four IOL calculation formulas between the four devices.

CONCLUSIONS:

The Anterion showed good repeatability of biometric measurements for most parameters. Good agreement among the four optical biometers was achieved for all the parameters except for CCT and the predicted IOL power. The AL values exhibited the best repeatability with Anterion and the best agreement among the biometers in our study.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Longitud Axial del Ojo / Lentes Intraoculares Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Graefes Arch Clin Exp Ophthalmol Año: 2022 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Longitud Axial del Ojo / Lentes Intraoculares Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Graefes Arch Clin Exp Ophthalmol Año: 2022 Tipo del documento: Article
...