In vitro and in vivo assessment of a novel ultra-flexible ventriculoamniotic shunt for treating fetal hydrocephalus.
Emery, Stephen P; Greene, Stephanie; Elsisy, Moataz; Chung, Kaitlin; Ye, Sang-Ho; Kim, Seungil; Wagner, William R; Hazen, Nika; Chun, Youngjae.
Afiliación
  • Emery SP; Department of Obstetrics, Gynecology & Reproductive Sciences, Divisions of Maternal-Fetal Medicine, 6620Magee Womens Hospital of UPMC, Pittsburgh, PA, USA.
  • Greene S; Department of Neurological Surgery, Division of Neurosurgery, 6619Children's Hospital of Pittsburgh of UPMC, Pittsburgh, PA, USA.
  • Elsisy M; Mechanical Design and Production Department, 63526Cairo University, Giza, Egypt.
  • Chung K; Department of Biengineering, 6614University of Pittsburgh, Pittsburgh, PA, USA.
  • Ye SH; Department of Surgery, 6595UPMC, Pittsburgh, PA, USA.
  • Kim S; Department of Surgery, 6595UPMC, Pittsburgh, PA, USA.
  • Wagner WR; Department of Surgery, 6595UPMC, Pittsburgh, PA, USA.
  • Hazen N; Center for Preclinical Studies, 536993University of Pittsburgh McGowan Institute for Regenerative Medicine, Pittsburgh, PA, USA.
  • Chun Y; Department of Biengineering, 6614University of Pittsburgh, Pittsburgh, PA, USA.
J Biomater Appl ; 37(8): 1423-1435, 2023 03.
Article en En | MEDLINE | ID: mdl-36063383
ABSTRACT
Fetal aqueductal stenosis (AS) is one of the most common causes of congenital hydrocephalus, which increases intracranial pressure due to partial or complete obstruction of cerebrospinal fluid (CSF) flow within the ventricular system. Approximately 2-4 infants per 10,000 births develop AS, which leads to progressive hydrocephalus, which enlarges the head often necessitating delivery by cesarean section. Most babies born with AS are severely neurologically impaired and experience a lifetime of disability. Therefore, a new device technology for venticuloamniotic shunting is urgently needed and has been studied to ameliorate or prevent fetal hydrocephalus development, which can provide a significant impact on patients and their family's quality of life and on the decrease of the healthcare dollars spent for the treatment. This study has successfully validated the design of shunt devices and demonstrated the mechanical performance and valve functions. A functional prototype shunt has been fabricated and subsequently used in multiple in vitro tests to demonstrate the performance of this newly developed ventriculoamniotic shunt. The shunt contains a main silicone-nitinol composite tube, a superelastic 90° angled dual dumbbell anchor, and an ePTFE valve encased by a stainless-steel cage. The anchor will change its diameter from 1.15 mm (collapsed state) to 2.75 mm (deployed state) showing up to 1.4-fold diameter change in human body temperature. Flow rates in shunts were quantified to demonstrate the valve function in low flow rates mimicking the fetal hydrocephalus condition showing "no backflow" for the valved shunt while there is up to 15 mL/h flow through the shunt with pressure difference of 20 Pa. In vivo ovine study results show the initial successful device delivery and flow drainage with sheep model.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cesárea / Hidrocefalia Aspecto: Patient_preference Límite: Animals / Female / Humans / Pregnancy Idioma: En Revista: J Biomater Appl Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cesárea / Hidrocefalia Aspecto: Patient_preference Límite: Animals / Female / Humans / Pregnancy Idioma: En Revista: J Biomater Appl Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos
...