Your browser doesn't support javascript.
loading
Developing a novel device, Eggcell, to improve temperature stability during oocyte collection for IVF.
Prathalingam, Nilendran; Hyslop, Louise; Cole, Michael; Cooney, Daniel; Driver, Alex; Herbert, Mary; Hodgson, Robert; Choudhary, Meenakshi; Wetherall, Ben; Shivhare, Sourima; Richardson, Jessica; Smith, Tom; Stewart, Jane; Murdoch, Alison.
Afiliación
  • Prathalingam N; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK. Electronic address: N.Prathalingam@nhs.scot.
  • Hyslop L; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Cole M; Population Health Sciences Institute, Newcastle University, Newcastle upon Tyne NE2 4AX, UK.
  • Cooney D; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Driver A; Labman Automation Ltd, Seamer Hill, Stokesley, North Yorkshire, TS9 5NQ, UK.
  • Herbert M; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Hodgson R; Labman Automation Ltd, Seamer Hill, Stokesley, North Yorkshire, TS9 5NQ, UK.
  • Choudhary M; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Wetherall B; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Shivhare S; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Richardson J; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Smith T; Labman Automation Ltd, Seamer Hill, Stokesley, North Yorkshire, TS9 5NQ, UK.
  • Stewart J; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
  • Murdoch A; Newcastle Fertility Centre at Life, Newcastle Hospitals NHS Trust, Newcastle upon Tyne, NE1 4EP, UK.
Reprod Biomed Online ; 45(6): 1097-1104, 2022 12.
Article en En | MEDLINE | ID: mdl-36207253
ABSTRACT
RESEARCH QUESTION What temperature fluctuations are oocytes exposed to during oocyte retrieval? Can an alternative method of oocyte retrieval be designed to minimize these fluctuations?

DESIGN:

Mock oocyte retrieval procedures were performed to investigate the change in temperature when the follicular fluid is drained into collection tubes and when the fluid is subsequently poured into dishes to allow identification of the cumulus-oocyte complex (COC). A new device, the Eggcell, has been designed that addresses the problem of these temperature fluctuations. To confirm its safety and demonstrate the clinical applicability of Eggcell, laboratory validation was performed prior to use with human participants (n = 15).

RESULTS:

Eggcell meets its design specification to provide temperature stability within the physiological range for aspirated follicular fluid. The COC can be successfully retained within the chamber (n = 180) without evidence of loss or damage to the oocytes or compromise of fertilization rate, blastocyst development or clinical outcome.

CONCLUSIONS:

This study has demonstrated the successful first stages of development of a new medical device. Further studies are needed for comparative evaluation of clinical outcome with standard technology.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fertilización In Vitro / Recuperación del Oocito Límite: Female / Humans Idioma: En Revista: Reprod Biomed Online Asunto de la revista: MEDICINA REPRODUTIVA Año: 2022 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fertilización In Vitro / Recuperación del Oocito Límite: Female / Humans Idioma: En Revista: Reprod Biomed Online Asunto de la revista: MEDICINA REPRODUTIVA Año: 2022 Tipo del documento: Article
...