Evaluation of x-ray effective focal spot size dependency on x-ray exposure settings using edge response analysis.
Shimakawa, Yurie; Nishiki, Masayuki; Yanagita, Satoshi; Nishikawa, Noriko; Sakai, Takayuki; Ito, Hajime; Ochi, Shigehiro; Yanagawa, Noriyuki.
Afiliación
  • Shimakawa Y; Department of Radiology, Eastern Chiba Medical Center, Togane, Japan.
  • Nishiki M; Department of Radiological Sciences, International University of Health and Welfare, Ohtawara, Japan. nishiki@iuhw.ac.jp.
  • Yanagita S; Department of Radiological Technology, Tsukuba International University, Tsuchiura, Japan.
  • Nishikawa N; Department of Radiological Sciences, International University of Health and Welfare, Ohtawara, Japan.
  • Sakai T; Department of Radiology, Eastern Chiba Medical Center, Togane, Japan.
  • Ito H; Department of Radiology, Eastern Chiba Medical Center, Togane, Japan.
  • Ochi S; Department of Radiology, Eastern Chiba Medical Center, Togane, Japan.
  • Yanagawa N; Department of Radiological Technology, Tsukuba International University, Tsuchiura, Japan.
Radiol Phys Technol ; 16(1): 39-47, 2023 Mar.
Article en En | MEDLINE | ID: mdl-36515865
ABSTRACT
The effective focal spot size of x-ray tubes is one of the major factors that substantially affect the resultant x-ray images, and it is known to be dependent on the x-ray exposure setting used. This study aims to evaluate the relationship between the effective focal spot size and the tube current and voltage and assess its reproducibility among several x-ray tubes. The evaluation was performed using edge response analysis, in which a 1-mm thick tungsten edge was projected onto a flat panel detector with a magnification factor of 2. The edge image was then differentiated to obtain the line spread function, followed by a detector blur-removing process through Fourier analysis to obtain the true focus profile. The resultant focal spot size increased as the tube current increased, whereas it decreased as the tube voltage increased, as expected. The rate of change was similar along the width and the length directions, while the small focus changed more significantly than the large focus. The reproducibility among four x-ray tubes of the same model was excellent as the maximum variation < 20%. In conclusion, the edge response method can provide useful information on the x-ray focal spot relationship with the x-ray exposure settings used, as well as its reproducibility among several x-ray tubes.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Intensificación de Imagen Radiográfica / Tomografía Computarizada por Rayos X Tipo de estudio: Prognostic_studies Idioma: En Revista: Radiol Phys Technol Asunto de la revista: BIOFISICA / RADIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Intensificación de Imagen Radiográfica / Tomografía Computarizada por Rayos X Tipo de estudio: Prognostic_studies Idioma: En Revista: Radiol Phys Technol Asunto de la revista: BIOFISICA / RADIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Japón
...