Establishment of Korean Pediatric Reference Intervals for Estradiol using Ultra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry.
Won, Eun Jeong; Yi, Ahram; Ko, Young Jin; Kim, Serim; Kang, Seong-Ho; Park, Geon; Jang, Sook-Jin.
Afiliación
  • Won EJ; Department of Laboratory Medicine, Chonnam National University Hwasun Hospital, Hwasun, South Korea.
  • Yi A; Department of Laboratory Medicine, Green Cross Laboratories (GC Labs), Yongin, Gyeonggi, South Korea.
  • Ko YJ; Department of Laboratory Medicine, College of Medicine, Chosun University, Gwangju, South Korea. Electronic address: yjko@chosun.ac.kr.
  • Kim S; Department of Laboratory Medicine, Green Cross Laboratories (GC Labs), Yongin, Gyeonggi, South Korea.
  • Kang SH; Department of Laboratory Medicine, College of Medicine, Chosun University, Gwangju, South Korea.
  • Park G; Department of Laboratory Medicine, College of Medicine, Chosun University, Gwangju, South Korea.
  • Jang SJ; Department of Laboratory Medicine, College of Medicine, Chosun University, Gwangju, South Korea.
Clin Biochem ; 113: 52-58, 2023 Mar.
Article en En | MEDLINE | ID: mdl-36627011
ABSTRACT

BACKGROUND:

Ultra-high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) is a reliable and accurate method for measuring steroid hormone levels. There is an increasing need for sensitive and precise methods to measure estradiol in pediatric patients. Here, we established reference intervals for estradiol in healthy children using a UHPLC-MS/MS-based method for the first time in South Korea.

METHODS:

Serum estradiol was measured using a Sciex Triple QuadTM 6500 + UHPLC-MS/MS (Sciex, Framingham, MA, USA). Reference intervals for estradiol were established according to the CLSI document EP28-A3c2008. The reference intervals were validated using serum samples from 634 pediatric patients, including neonates, children, and adolescents. Among them, 389 specimens were used in analysis of the specimen acceptance time. Statistical analysis was performed using MedCalc (MedCalc, Ostend, Belgium) and Analyse-it (Analyse-it Software Ltd., Leeds, United Kingdom) software.

RESULTS:

Reference intervals for boys (n = 297) were <16.6, <7.3, <19.0, <30.5, 7.6-96.5, and 10.6-134.4 pmol/L among those aged <1, 1-5, 6-9, 10-11, 12-14, and 15-17 years, respectively. Reference intervals for girls (n = 337) were <114.7, <24.2, <34.8, 8.0-177.0, 10.4-480.5, and 9.1-486.7 pmol/L among those aged <1, 1-5, 6-9, 10-11, 12-14, and 15-17 years, respectively. Overall, there was no effect of specimen acceptance time on estradiol measurements in boys or girls, except for that in the group aged 10-11 years.

CONCLUSIONS:

The reference intervals for healthy children were validated using a UHPLC-MS/MS-based method. The highly analytical sensitive UHPLC-MS/MS method may be useful for estradiol determination in pediatric patients.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Estradiol / Espectrometría de Masas en Tándem Límite: Adolescent / Child / Female / Humans / Male / Newborn Idioma: En Revista: Clin Biochem Año: 2023 Tipo del documento: Article País de afiliación: Corea del Sur

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Estradiol / Espectrometría de Masas en Tándem Límite: Adolescent / Child / Female / Humans / Male / Newborn Idioma: En Revista: Clin Biochem Año: 2023 Tipo del documento: Article País de afiliación: Corea del Sur
...