Comparing Remote Speckle Plethysmography and Finger-Clip Photoplethysmography with Non-Invasive Finger Arterial Pressure Pulse Waves, Regarding Morphology and Arrival Time.
Herranz Olazabal, Jorge; Wieringa, Fokko; Hermeling, Evelien; Van Hoof, Chris.
Afiliación
  • Herranz Olazabal J; IMEC, 3000 Leuven, Belgium.
  • Wieringa F; Faculty of Engineering Science, Katholieke Universiteit Leuven (KUL), 3000 Leuven, Belgium.
  • Hermeling E; IMEC NL, 5656 AE Eindhoven, The Netherlands.
  • Van Hoof C; IMEC NL, 5656 AE Eindhoven, The Netherlands.
Bioengineering (Basel) ; 10(1)2023 Jan 11.
Article en En | MEDLINE | ID: mdl-36671673
ABSTRACT

OBJECTIVE:

The goal was to compare Speckle plethysmography (SPG) and Photoplethysmography (PPG) with non-invasive finger Arterial Pressure (fiAP) regarding Pulse Wave Morphology (PWM) and Pulse Arrival Time (PAT).

METHODS:

Healthy volunteers (n = 8) were connected to a Non-Invasive Blood Pressure (NIBP) monitor providing fiAP pulse wave and PPG from a clinical transmission-mode SpO2 finger clip. Biopac recorded 3-lead ECG. A camera placed at a 25 cm distance recorded a video stream (100 fps) of a finger illuminated by a laser diode at 639 nm. A chest belt (Polar) monitored respiration. All signals were recorded simultaneously during episodes of spontaneous breathing and paced breathing.

ANALYSIS:

Post-processing was performed in Matlab to obtain SPG and analyze the SPG, PPG and fiAP mean absolute deviations (MADs) on PWM, plus PAT modulation.

RESULTS:

Across 2599 beats, the average fiAP MAD with PPG was 0.17 (0-1) and with SPG 0.09 (0-1). PAT derived from ECG-fiAP correlated as follows 0.65 for ECG-SPG and 0.67 for ECG-PPG.

CONCLUSION:

Compared to the clinical NIBP monitor fiAP reference, PWM from an experimental camera-derived non-contact reflective-mode SPG setup resembled fiAP significantly better than PPG from a simultaneously recorded clinical transmission-mode finger clip. For PAT values, no significant difference was found between ECG-SPG and ECG-PPG compared to ECG-fiAP.