Performance evaluation of the PennPET explorer with expanded axial coverage.
Dai, Bing; Daube-Witherspoon, Margaret E; McDonald, Stephen; Werner, Matthew E; Parma, Michael J; Geagan, Michael J; Viswanath, Varsha; Karp, Joel S.
Afiliación
  • Dai B; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
  • Daube-Witherspoon ME; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
  • McDonald S; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
  • Werner ME; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
  • Parma MJ; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
  • Geagan MJ; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
  • Viswanath V; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
  • Karp JS; Department of Radiology, University of Pennsylvania, Philadelphia, United States of America.
Phys Med Biol ; 68(9)2023 04 19.
Article en En | MEDLINE | ID: mdl-36958051
ABSTRACT
Objective.This work evaluated the updated PennPET Explorer total-body (TB) PET scanner, which was extended to 6 rings with updated readout firmware to achieve a 142 cm axial field of view (AFOV) without 7.6 cm inter-ring axial gaps.Approach.National Electrical Manufacturers Association (NEMA) NU 2-2018 measurements were performed with modifications including longer phantoms for sensitivity and count-rate measurements and additional positions for spatial resolution and image quality. A long uniform phantom and the clinical trials network (CTN) phantom were also used.Main results.The total sensitivity increased to 140 kcps MBq-1for a 70 cm line, a gain of 1.8x compared to the same system with axial gaps; an additional 47% increase in total counts was observed with a 142 cm line at the same activity per cm. The noise equivalent count rate (NECR) increased by 1.8x without axial gaps. The peak NECR is 1550 kcps at 25 kBq cc-1for a 140 cm phantom; due to increased randoms, the NECR is lower than with a 70 cm phantom, for which NECR is 2156 kcps cc-1at 25 kBq cc-1and continues increasing. The time-of-flight resolution is 250 ps, increasing by <10 ps at the highest activity. The axial spatial resolution degrades by 0.6 mm near the center of the AFOV, compared to 4 mm resolution near the end. The NEMA image quality phantom showed consistent contrast recovery throughout the AFOV. A long uniform phantom demonstrated axial uniformity of uptake and noise, and the CTN phantom demonstrated quantitative accuracy for both18F and89Zr.Significance. The performance evaluation of the updated PennPET Explorer demonstrates significant gains compared to conventional scanners and shows where the current NEMA standard needs to be updated for TB-PET systems. The comparisons of systems with and without inter-ring gaps demonstrate the performance trade-offs of a more cost-effective TB-PET system with incomplete detector coverage.