Technical and clinical assessment of latest technology SiPM integrated digital PETCT scanner.
Singh, M K; Mohan, P; Mahajan, H; Kaushik, C.
Afiliación
  • Singh MK; Medikabazaar, Technopolis Knowledge Park, Mumbai, 400093, India.
  • Mohan P; Mahajan Imaging, Hauz Khas Enclave, New Delhi, 110016, India.
  • Mahajan H; Mahajan Imaging, Hauz Khas Enclave, New Delhi, 110016, India.
  • Kaushik C; School of Health and Society, University of Salford, Manchester, M5 4WT, United Kingdom. Electronic address: C.Kaushik@salford.ac.uk.
Radiography (Lond) ; 29(4): 705-711, 2023 07.
Article en En | MEDLINE | ID: mdl-37187068
ABSTRACT

OBJECTIVE:

The aim of this study was to conduct a technical and clinical evaluation of a Silicon Photomultiplier (SiPM) integrated digital Positron Emission Tomography - Computed Tomography (PETCT) Scanner using National Electrical Manufacturers Association (NEMA) NU 2- 2018 standards.

METHODS:

System sensitivity was measured by using a NEMA sensitivity phantom. Scatter fraction, count-rate performance, accuracy of count loss, and timing resolution were all computed. Clinical images were acquired and image quality was assessed and compared with published studies.

RESULTS:

At 1 cm, tangential, radial, and axial spatial resolutions were 3.02 mm, 3.02 mm, and 2.73 mm at full width half maximum (FWHM), respectively. Sensitivity at centre and 10 cm was 10.359 cps/kBq and 9.741 cps/kBq, respectively. The timing resolution was measured at 372 ps.

CONCLUSION:

The digital PETCT exhibits a high-spatial resolution and a superior timing resolution, which advances the diagnostic ability to detect small lesions and boosts the diagnostic confidence. IMPLICATIONS FOR PRACTICE Increases clinical relevance by improving the ability to detect and differentiate tiny or low-contrast lesions without compromising radiopharmaceutical dose or overall scan time.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía de Emisión de Positrones / Tomografía Computarizada por Tomografía de Emisión de Positrones Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: Radiography (Lond) Año: 2023 Tipo del documento: Article País de afiliación: India

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía de Emisión de Positrones / Tomografía Computarizada por Tomografía de Emisión de Positrones Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: Radiography (Lond) Año: 2023 Tipo del documento: Article País de afiliación: India
...