A method for selecting reference beam model of VMAT plans with three 6MV beam-matched linear accelerators during radiation oncology.
Li, Yi; Wu, Wenjing; Yuan, Wei; Chai, Linyan; Tang, Fengwen; He, Ruixin; Lu, Yongkai; Zhang, Yuemei; Lu, Yongkai; Wang, Long; Wang, Mengfei; Zhang, Xiaozhi; Zhang, Long.
Afiliación
  • Li Y; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Wu W; Department of Radiological Health, Xi'an Center for Disease Control and Prevention, Xi'an, 710054, China. wu_ermao@163.com.
  • Yuan W; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Chai L; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Tang F; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • He R; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Lu Y; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Zhang Y; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Lu Y; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Wang L; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Wang M; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Zhang X; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China.
  • Zhang L; Department of Radiation Oncology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, 710061, China. zhanglonglongmail@163.com.
Sci Rep ; 13(1): 10131, 2023 06 22.
Article en En | MEDLINE | ID: mdl-37349531
ABSTRACT
Our objective was to provide a method for selecting reference beam model and evaluating the dosimetric accuracy of volumetric modulated arc therapy (VMAT) plans delivered on three Elekta beam-matched linacs during radiation oncology. Beam data was measured on three beam-matched linacs including Synergy1, Synergy2 and VersaHD. For eighteen lung and esophagus cases, fifty-four plans were generated using VMAT technique with three linac beam models respectively for point dose measurement and three-dimensional dose measurement. Each VMAT plan was executed sequentially on three linacs respectively. Measurement results were compared with treatment planning system (TPS) calculation results for all VMAT plans. Among three beam-matched linacs, discrepancy in beam output factor, percentage depth dose at 5 cm, 10 cm, 20 cm depth and MLC leaf offset are all within 1% except 20 × 20 cm2 and 30 × 30 cm2 field sizes, and discrepancy in beam profile is all within 2%. With comparison between measurement result and TPS calculation result, the absolute dose deviations are within the range of ± 3%, and the gamma passing rates are all over 95% for all VMAT plans, which are within the tolerance of clinical acceptability. Compared with all plans delivered on Synegy1 and VersaHD, the point dose discrepancy between measured results and TPS calculated results for plans delivered on Synergy2 is smallest, and the gamma passing rate between measured results and TPS calculated results for plans delivered on Synergy2 is highest. The beam-matched linacs demonstrate good agreement between measurement result and TPS calculation result for VMAT plans. The method can be used for selecting reference beam model for VMAT plans.