The effects of different metal posts, cements, and exposure parameters on cone-beam computed tomography artifacts.
Freitas, Ana Priscila Lira de Farias; Peixoto, Larissa Rangel; Mariz Suassuna, Fernanda Clotilde; Bento, Patrícia Meira; Maia Amorim, Ana Marly Araújo; Rovaris Silva, Karla; de Almeida Barros, Renata Quirino; Pontual de Andrade Lima, Andrea Dos Anjos; de Melo, Daniela Pita.
Afiliación
  • Freitas APLF; Department of Oral Diagnosis, State University of Paraíba, Campina Grande, PB, Brazil.
  • Peixoto LR; Department of Oral Diagnosis, State University of Paraíba, Campina Grande, PB, Brazil.
  • Mariz Suassuna FC; Department of Oral Diagnosis, State University of Paraíba, Campina Grande, PB, Brazil.
  • Bento PM; Department of Oral Diagnosis, State University of Paraíba, Campina Grande, PB, Brazil.
  • Maia Amorim AMA; Department of Oral Diagnosis, State University of Paraíba, Campina Grande, PB, Brazil.
  • Rovaris Silva K; Department of Oral Diagnosis, Federal University of Piauí, Teresina, PI, Brazil.
  • de Almeida Barros RQ; Department of Oral Imaging of Unifacisa, Campina Grande, PB, Brazil.
  • Pontual de Andrade Lima ADA; Department of Oral Diagnosis, Federal University of Pernambuco, Campina Grande, PB, Brazil.
  • de Melo DP; Department of Oral Diagnosis, State University of Paraíba, Campina Grande, PB, Brazil.
Imaging Sci Dent ; 53(2): 127-135, 2023 Jun.
Article en En | MEDLINE | ID: mdl-37405202
ABSTRACT

Purpose:

This study assessed the intensity of artifacts produced by 2 metal posts, 2 cements, and different exposure parameters using 2 cone-beam computed tomography (CBCT) units. Materials and

Methods:

The sample was composed of 20 single-rooted premolars, divided into 4 groups Ni-Cr/zinc phosphate, Ni-Cr/resin cement, Ag-Pd/zinc phosphate, and Ag-Pd/resin cement. Samples were scanned before and after post insertion and cementation using a CS9000 3D scanner with 4 exposure parameters (85/90 kV and 6.3/10 mA) and an i-CAT scanner with 120 kV and 5 mA. The presence of artifacts was assessed subjectively by 2 observers and objectively by a trained observer using ImageJ software. The Mann-Whitney, Wilcoxon, weighted kappa, and chi-square tests were used to assess data at a 95% confidence level (α<0.05).

Results:

In the subjective analyses, AgPd presented more hypodense and hyperdense lines than NiCr (P<0.05), and more hypodense halos were found using i-CAT (P<0.05) than using CS9000 3D. More hypodense halos, hypodense lines, and hyperdense lines were observed at 10 mA than at 6.3 mA (P<0.05). More hypodense halos were observed at 85 kV than at 90 kV (P<0.05). CS9000 3D presented more hypodense and hyperdense lines than i-CAT (P<0.05). In the objective analyses, AgPd presented higher percentages of hyperdense and hypodense artifacts than NiCr (P<0.05). Zinc phosphate cement presented higher hyperdense artifact percentages on CS9000 3D scans (P<0.05). CS9000 3D presented higher artifact percentages than i-CAT (P<0.05).

Conclusion:

High-atomic-number alloys, higher tube current, and lower tube voltage may increase the artifacts present in CBCT images.