Initial description of the novel handheld wireless ultrasound device TE Air with Doppler and Color Duplex imaging.
Jung, Ernst Michael; Jung, Friedrich; Dong, Yi; Kaiser, Ulrich.
Afiliación
  • Jung EM; Institute for Diagnostic Radiology and Interdisciplinary Ultrasound, University Hospital Regensburg, Regensburg, Germany.
  • Jung F; Institute of Biotechnology, Molecular Cell Biology, Brandenburg University of Technology, Senftenberg, Germany.
  • Dong Y; Xinhua Hospital affiliated to Shanghai Jiaotong University School of Medicine, Shanghai, China.
  • Kaiser U; Medical Clinic and Polyclinic III, University Hospital Regensburg, Regensburg, Germany.
Clin Hemorheol Microcirc ; 86(1-2): 89-97, 2024.
Article en En | MEDLINE | ID: mdl-37574725
ABSTRACT

OBJECTIVE:

To test and initially describe a new handheld wireless ultrasound technique (TE Air) for clinical use.

METHODS:

In this pilot study, the new ultrasound device TE Air from Mindray was used to examine the hepatic and renal vessels of healthy volunteers for first impressions. The probe has a sector transducer with a frequency range of 1.8-4.5 MHz. The B-mode and color-coded doppler sonography (CCDS) scanning methods were used. A high-end device from the same company (Resona 9, Mindray) was used as a reference. The results were evaluated using an image rating scale ranging from 0 to 5, with 0 indicating not assessable and 5 indicating without limitations.

RESULTS:

Altogether, 61 participants (n = 34 female [55.7%], n = 27 male [44.3%]), age range 18-83 years, mean age 37.9±16.5 years) could be adequately studied using TE AIR and the high-end device. With one exception, the image quality score for TE Air never fell below 3 and had a mean/median scored of 4.97/5.00 for the B-mode, 4.92/5.00 for the color flow (CF) mode, and 4.89/5.00 for the pulse wave (PW) mode of the hepatic vein, 4.90/5.00 for the portal vein, 4.11/4.00 for the hepatic artery, and 4.57/5.00 for the renal segmental artery. A significant difference in the assessment of flow measurement of the hepatic artery and renal segmental arteries was found between TE AIR and the high-end device.

CONCLUSIONS:

TE Air represents a new dimension in point-of-care ultrasound via wireless handheld devices. Especially, its flow measurement ability offers a relevant advantage over other available handheld models. TE Air provides a formally sufficient image quality in terms of diagnostic significance.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Vena Porta / Ultrasonografía Doppler en Color Tipo de estudio: Prognostic_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Hemorheol Microcirc Asunto de la revista: ANGIOLOGIA / HEMATOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Vena Porta / Ultrasonografía Doppler en Color Tipo de estudio: Prognostic_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Hemorheol Microcirc Asunto de la revista: ANGIOLOGIA / HEMATOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Alemania
...