Absorbed dose evaluation of a blood irradiator with alanine, TLD-100 and ionization chamber.
Appl Radiat Isot
; 200: 110981, 2023 Oct.
Article
en En
| MEDLINE
| ID: mdl-37633189
Irradiation of blood bags using X-ray irradiators and dosimetry services are required to ensure uniform dose levels in the range 25-50 Gy to prevent Transfusion Associated Graft versus Host Disease (TA-GvHD). An absorbed dose characterization of a Raycell MK2 X-Irradiator was performed using three different dosimetric systems. Results showed a dosimetric accuracy of the ionization chamber together with the Alanine dosimeter. TLDs measurements exhibited a small overestimation by 4% of the absorbed dose. The Dose Uniformity Ratio (DUR), between maximum and minimum dose levels in the canister, was in good agreement with the manufacturer specifications (≤1.5).
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Alanina
/
Dosímetros de Radiación
Idioma:
En
Revista:
Appl Radiat Isot
Asunto de la revista:
MEDICINA NUCLEAR
/
SAUDE AMBIENTAL
Año:
2023
Tipo del documento:
Article
País de afiliación:
Italia