Your browser doesn't support javascript.
loading
A computer vision based optical method for measuring fluid level in cell culture plates.
Baudin, Pierre V; Teodorescu, Mircea.
Afiliación
  • Baudin PV; Department of Electrical and Computer Engineering, University of California Santa Cruz, Santa Cruz, CA, United States of America.
  • Teodorescu M; Genomics Institute, University of California Santa Cruz, Santa Cruz, CA, United States of America.
PLoS One ; 18(9): e0290951, 2023.
Article en En | MEDLINE | ID: mdl-37682933
ABSTRACT
For a transparent well with a known volume capacity, changes in fluid level result in predictable changes in magnification of an overhead light source. For a given well size and fluid, the relationship between volume and magnification can be calculated if the fluid's index of refraction is known or in a naive fashion with a calibration procedure. Light source magnification can be measured through a camera and processed using computer vision contour analysis with OpenCV. This principle was applied in the design of a 3D printable sensing device using a raspberry pi zero and a camera.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Refracción Ocular / Técnicas de Cultivo de Célula Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos