Effects of Steady Glucose Concentration Peritoneal Dialysis on Ultrafiltration Volume and Sodium Removal: A Pilot Crossover Trial.
Heimbürger, Olof; Hegbrant, Jörgen; Martus, Giedre; Wilkie, Martin; De Leon, Charlotte; Carlsson, Ola; Johansson, Ann-Cathrine.
Afiliación
  • Heimbürger O; Medical Unit Renal Medicine, Karolinska University Hospital, and CLINTEC, Karolinska Institutet, Stockholm, Sweden.
  • Hegbrant J; Division of Nephrology, Department of Clinical Sciences, Lund University, Lund, Sweden.
  • Martus G; Department of Nephrology, Skåne University Hospital, Lund, Sweden.
  • Wilkie M; Sheffield Teaching Hospitals, Sheffield, United Kingdom.
  • De Leon C; Triomed AB, Lund, Sweden.
  • Carlsson O; Triomed AB, Lund, Sweden.
  • Johansson AC; Department of Nephrology, Skåne University Hospital, Malmö, Sweden.
Article en En | MEDLINE | ID: mdl-37902732
ABSTRACT

BACKGROUND:

Volume overload is common in patients treated with peritoneal dialysis (PD) and is associated with poor clinical outcome. Steady concentration PD is where a continuous glucose infusion maintains the intraperitoneal glucose concentration and as a result provides continuous ultrafiltration throughout the dwell. The primary objective of this study was to investigate the ultrafiltration rate and glucose ultrafiltration efficiency for steady concentration PD in comparison with a standard continuous ambulatory PD (CAPD) dwell, using the novel Carry Life UF device.

METHODS:

Eight stable patients treated with PD (six fast and two fast average transporters) were investigated four times a standard 4-hour CAPD dwell with 2 L of 2.5% dextrose solution as control and three 5-hour steady concentration PD treatments (glucose dose 11, 14, 20 g/h, initial fill 1.5 L of 1.5% dextrose solution). All investigations were preceded by an overnight 2 L 7.5% icodextrin dwell.

RESULTS:

Intraperitoneal glucose concentration increased during the first 1-2 hours of the steady concentration PD treatments and remained stable thereafter. Ultrafiltration rates were significantly higher with steady concentration PD treatments (124±49, 146±63, and 168±78 mL/h with 11, 14, and 20 g/h, respectively, versus 40±60 mL/h with the control dwell). Sodium removal and glucose ultrafiltration efficiency (ultrafiltration volume/gram glucose uptake) were significantly higher with steady concentration PD treatments versus the control dwell, where the 11 g/h glucose dose was most efficient.

CONCLUSIONS:

Steady concentration PD performed with the Carry Life UF device resulted in higher ultrafiltration rates, more efficient use of glucose (increased ultrafiltration volume/gram glucose absorbed), and greater sodium removal compared with a standard 2.5% dextrose CAPD dwell. CLINICAL TRIAL REGISTRY NAME AND REGISTRATION NUMBER A Performance Analysis of the Peritoneal Ultrafiltration (PUF) Achieved With the Carry Life ® UF, NCT03724682 .