Your browser doesn't support javascript.
loading
Effects on the accelerating electron bunches due to the presence of sulfur hexafluoride or air in the linac waveguide.
Gayol, A; Figueroa, R; Guarda, J; Leiva, J; Leyton, F; Malano, F; Valente, M.
Afiliación
  • Gayol A; Instituto de Física Enrique Gaviola, CONICET & Laboratorio de Investigaciones e Instrumentación en Física Aplicada a La Medicina e Imágenes por Rayos X - LIIFAMIRx, FAMAF, Universidad Nacional de Córdoba, Córdoba, 5000, Argentina.
  • Figueroa R; Centro de Física e Ingeniería en Salud CFIS, Universidad de La Frontera, Temuco, 4780000, Chile; Departamento de Ciencias Físicas, Universidad de La Frontera, Temuco, 4780000, Chile.
  • Guarda J; Centro de Física e Ingeniería en Salud CFIS, Universidad de La Frontera, Temuco, 4780000, Chile.
  • Leiva J; Centro de Física e Ingeniería en Salud CFIS, Universidad de La Frontera, Temuco, 4780000, Chile.
  • Leyton F; Centro de Física e Ingeniería en Salud CFIS, Universidad de La Frontera, Temuco, 4780000, Chile.
  • Malano F; Centro de Física e Ingeniería en Salud CFIS, Universidad de La Frontera, Temuco, 4780000, Chile; Departamento de Ciencias Físicas, Universidad de La Frontera, Temuco, 4780000, Chile.
  • Valente M; Instituto de Física Enrique Gaviola, CONICET & Laboratorio de Investigaciones e Instrumentación en Física Aplicada a La Medicina e Imágenes por Rayos X - LIIFAMIRx, FAMAF, Universidad Nacional de Córdoba, Córdoba, 5000, Argentina; Centro de Física e Ingeniería en Salud CFIS, Universidad de La Fr
Appl Radiat Isot ; 205: 111155, 2024 Mar.
Article en En | MEDLINE | ID: mdl-38159451
ABSTRACT
Sulfur hexafluoride gas (SF6) is used as a dielectric insulator in the acceleration process of certain medical linear accelerator waveguides. Nevertheless, some innovative development and investigation cases require intervention in the linear accelerator or, specifically, on the waveguide, which could affect the sealing of the device. In this regard, vacuum sealing systems can be compromised, affecting the properties of the radiation beams produced. The presence of sulfur hexafluoride or air inside the VARIAN 6/100 waveguide was investigated under different pressure conditions and non-uniform electric fields, adapting Monte Carlo simulation techniques for modeling radiation transport coupled with electric fields. Obtained results indicated the suitability of the proposed approach, while comparisons with theoretical approaches and experimental evidence supported the model's consistency.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2024 Tipo del documento: Article País de afiliación: Argentina

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Appl Radiat Isot Asunto de la revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Año: 2024 Tipo del documento: Article País de afiliación: Argentina
...