A case of endocardial dissection caused by Micra implantation.
Zhang, Jianghua; Shang, Shuai; Dong, Zhenyu; Zhou, Xianhui; Li, Yaodong; Lu, Yanmei; Xing, Qiang; Tuerhong, Zukela; Guo, Yankai; Xiaokereti, Jiasuoer; Tang, Baopeng.
Afiliación
  • Zhang J; Department of Pacing and Electrophysiology, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Shang S; Xinjiang Key Laboratory of Cardiac Electrophysiology and Remodeling, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Dong Z; Department of Pacing and Electrophysiology, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Zhou X; Xinjiang Key Laboratory of Cardiac Electrophysiology and Remodeling, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Li Y; Department of Pacing and Electrophysiology, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Lu Y; Xinjiang Key Laboratory of Cardiac Electrophysiology and Remodeling, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Xing Q; Department of Pacing and Electrophysiology, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Tuerhong Z; Xinjiang Key Laboratory of Cardiac Electrophysiology and Remodeling, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Guo Y; Department of Pacing and Electrophysiology, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Xiaokereti J; Xinjiang Key Laboratory of Cardiac Electrophysiology and Remodeling, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
  • Tang B; Department of Pacing and Electrophysiology, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.
BMC Cardiovasc Disord ; 24(1): 10, 2024 01 02.
Article en En | MEDLINE | ID: mdl-38166609
ABSTRACT

BACKGROUND:

Leadless pacemakers are a recent technological advancement. It has many advantages, but there are still a few serious complications. CASE PRESENTATION This article reports the case of a patient with an endocardial tear and dissection caused by contact with the tip of the Micra cup during surgery and summarises the relevant data.

CONCLUSIONS:

This case report details the occurrence and management of the incident and provides some guidance for future clinical management.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: BMC Cardiovasc Disord Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: BMC Cardiovasc Disord Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: China
...